Inferior Vena Cava Filters 

  • Author: Gary P Siskin, MD; Chief Editor: Kyung J Cho, MD, FACR, FSIR  more...
 
Updated: Dec 03, 2015
 

Products

Inferior vena cava (IVC) filter placement is most commonly indicated for deep venous thrombosis (DVT) or pulmonary embolism (PE) when anticoagulation therapy is contraindicated.

The first image below demonstrates an IVC of normal size without a thrombus. The second image below shows on type of IVC filter in place.

Anteroposterior image from an inferior venacavogra Anteroposterior image from an inferior venacavographic examination demonstrates an inferior vena cava of normal diameter without thrombus. In addition, the inflow from the renal veins can be seen at the level of the L1 vertebral body.
Greenfield filter. Greenfield filter.

Category

Inferior vena cava filters

Device details

Permanent filters

Bard Peripheral Vascular

  • Simon Nitinol Filter (see the following image)
    Simon nitinol filter. Simon nitinol filter.

B Braun Medical

  • Vena Tech LGM vena cava filter
  • Vena Tech LP vena cava filter

Boston Scientific

  • Stainless steel Greenfield filter (SGF) (Kimray-Greenfield filter)
  • Titanium Greenfield filter (TGF)
  • Percutaneous stainless steel Greenfield filter (12F SGF)
  • An example of an inserted Greenfield filter is shown in the image below
    Greenfield filter. Greenfield filter.

Cook Medical

  • Gianturco-Roehm Bird's Nest filter (BNF) (see the following image)
    Bird's nest filter. Bird's nest filter.

Cordis

  • TrapEase permanent vena cava filter (see the image below)
    TrapEase filter. TrapEase filter.

Rafael Medical Technologies

  • SafeFlo vena cava filter

Retrievable/optional filters

ALN Implants Chirugicaux

  • ALN optional vena cava filter

Bard Peripheral Vascular

  • Recovery vena cava filter (discontinued)
  • G2 vena cava filter
  • G2 X vena cava filter
  • Eclipse vena cava filter

Cook Medical

  • Günther Tulip vena cava filter
  • Celect vena cava filter

Cordis

  • OptEase retrievable vena cava filter

Rex Medical

  • Option retrievable vena cava filter (previously licensed by Angiotech; currently licensed by Argon Medical Devices)
Next

Design Features

General considerations

Desired quality of IVC filters

Inferior vena cava (IVC) filters are designed for their physical properties, clot-trapping effectiveness, ability to preserve flow in the IVC, and ease of placement. Of the currently available filters, each type has some but not all of the following properties or capabilities:

  • The filter should be able to trap most, if not all, thrombi to prevent new or recurrent pulmonary emboli (PE)
  • The filter should be nonthrombogenic and should maintain caval patency
  • The filter should be made of a biocompatible material that is durable and noncorrosive
  • The filter's shape and structural integrity should be maintainable for a long time
  • The filter's delivery system should have a low profile and allow easy placement
  • Clot trapping should be reasonably effective even if the filter deployment is suboptimal
  • The filter should not migrate after deployment
  • No perforation of the IVC should occur
  • The filter should be nonferromagnetic to allow magnetic resonance imaging (MRI) to be performed after its placement
  • The filter may be retrievable

Technical considerations of filter placement

Mobin-Uddin umbrella and 24F Kimray-Greenfield filters were originally designed for placement through a surgical venotomy. All of the other currently available filters can be placed percutaneously.

According to manufacturer recommendations, certain filters cannot or should not be placed by using certain approaches. When filter design allows placement through the jugular or femoral veins, the filter is specially packaged to ensure that it is deployed in the correct orientation.

Most filter placement procedures are performed in the angiographic suite with fluoroscopic guidance. In recent years, IVC filters have been placed successfully at the bedside in trauma patients in critical condition.[1] The procedures may be performed under guidance with a C-arm portable radiographic unit or with an intravascular ultrasonographic (US) scanner.

Most filters are placed in the infrarenal IVC. On the basis of the knowledge gained from caval ligation, some investigators advocate placement of the apex of any filter specifically within the inflow from the lowest renal vein to minimize the dead space between the filter and the renal veins. Suprarenal filter placement, when necessary, has been proven to be safe. Placement of a suprarenal IVC filter may be indicated in the following situations:

  • Renal vein thrombosis
  • IVC thrombosis extending above the level of the renal veins
  • Thrombus in the infrarenal IVC that does not leave enough room for the filter between the thrombus and renal veins
  • Recurrent PE despite infrarenal filter placement (in which upper extremity or superior vena cava [SVC] thrombosis should be excluded)
  • PE after ovarian vein thrombosis
  • Pregnancy

Regarding SVC filter placement, upper-extremity deep vein thrombosis (DVT) accounts for fewer than 5% of cases of venous thrombosis. The incidence has increased since the 1960s because of the more common use of indwelling catheters. Central venous catheters are responsible for 17-47% of reported upper-extremity DVTs. PE develops in as many as 12% of patients with upper-extremity venous thrombosis. Treatment options include rest and elevation of the affected arm, anticoagulation therapy, and thrombolysis. Filters have been deployed in the SVC and can effectively prevent symptomatic or fatal PE.

Placement of an SVC filter requires use of a modified technique. To maintain the correct orientation of the filter, a filter designed for jugular access is placed via a femoral approach, whereas the filter designed for femoral access is placed via a jugular approach.

It is more technically demanding to perform percutaneous insertion of an SVC filter than placement of an IVC filter, because the SVC is shorter and has a relatively smaller area for safe filter deployment. However, in reported series, no evidence of filter migration, dislodgement, or fracture was observed.

IVC filters

Mobin-Uddin umbrella filter

The Mobin-Uddin umbrella filter was the first true intravascular device and is included here for historical interest. This filter was discontinued in 1986.

Greenfield filter

In 1973, Greenfield et al introduced a new stainless steel filter.[2] The Greenfield filter has been on the market the longest, it is the most extensively evaluated filter, and it has become the criterion standard with which other IVC filters are compared. (See the image below.)

Greenfield filter. Greenfield filter.

The original Greenfield filter is conical, and it consists of 6 strands of 0.015-in, zigzag-shaped, 316L-grade stainless steel legs in a radial array, each with a hook at the end that anchors the filter to the IVC. The steel struts are 4.6 cm in length and are fitted to an apical hub, which is directed cephalad in the patient. The base is 3.0 cm in diameter with the legs separated by 11 mm in the deployed position. The filter is loaded into a 24F carrier with a 29.5F external diameter. Initially, the filter was placed via a surgical cutdown of either the right internal jugular vein or the right common femoral vein. Tadavarthy et al were the first to describe percutaneous placement of the Greenfield filter in 1984.[3]

The design of the Greenfield filter made it possible for thrombi to fill and occlude 70% of the filter cone, representing a volume of thrombus of approximately 4 cm3 (34.3% of total volume) without reducing the cross-sectional area by 50%. If the cone is filled to 80% of total volume, the reduction of cross-sectional area is 64%. The ability of this filter to trap clinically significant emboli has been demonstrated in several in vitro and in vivo studies.

In 1995, Greenfield et al reported on their 20-year experience with the Greenfield filter.[4] Recurrent PE was seen in 4% of 642 filter placements. The rate was comparable to that reported by other authors. Athanasoulis et al reported an overall recurrent PE rate of 4.9% in their experience with 7 different designs.[5] In the group of patients receiving the standard 24F Greenfield filter, 38 (8.4%) of 455 patients had recurrent PE.[5] Kantor et al reported access site thrombosis in 7 (41%) of 17 patients; this rate of thrombosis was associated with the large size of the delivery system.[6]

In 101 patients evaluated, Greenfield et al reported patency of the insertion site in 94 patients (93%).[4] In the same series, the authors reported an IVC patency rate of 96%. Migration of the filter was reported in 30-49% of patients. Allowing for respiratory variation, Greenfield et al readjusted the rate of migration of more than 3 mm to 8%.[4] Filter tilt, crossing of limbs, and limb fracture occur in a low percentage of patients, and most of these events are not associated with clinical sequelae.

Extension of the filter struts beyond the IVC, which is suggestive of perforation, has been noted in high incidence on computed tomography (CT) scanning. Proctor et al performed animal studies to evaluate the apparent vena caval penetration by the Greenfield filter.[7] All filters appeared to penetrate the IVC on cavography and CT scanning. Laparoscopy failed to demonstrate penetration, and histology revealed remodeling of the intimal surface of the IVC and thinning of the adventitia.[7] The authors further hypothesized that the presence of the filter caused adaptation and remodeling of the IVC to occur.

Titanium Greenfield filter

The titanium Greenfield filter (TGF) was introduced in 1989 as a low-profile system to facilitate the ease of percutaneous insertion. It is made of a titanium alloy with elastic properties that allow the filter to be fitted into a 12F carrier system, as opposed to the 24F stainless steel Greenfield filter (SGF). The design of the filter preserves the conical shape consisting of 6 radiating, zigzag legs.

The TGF measures 47 mm; by contrast, the SGF measures 46 mm. The base of the TGF is 38 mm, as compared with 30 mm in the SGF. At 0.25 g, the TGF is lighter than the SGF, which weighs 0.56 g. The material is resistant to flexion fatigue and corrosion, and is nonthrombogenic and nonferromagnetic. The filter can be inserted via either the right or left femoral or jugular vein through a dilator system with an attached 14F sheath. The femoral and the jugular systems are supplied with differently coded packages.

During their preliminary clinical studies, Greenfield et al found an unacceptable 30% rate of migration, tilting, and penetration.[8] The TGF has greater lateral force at the base than the SGF, which may result in a higher rate of IVC penetration. The hooks have been modified through development and evolution.[9] Experience with the TGF with modified hooks (TGF-MH) was reported in 1991. Migration was greater than 9 mm in 11% of patients.[10] There was an increase in base diameter by greater than 5 mm in 14% of patients (with 1 filter penetration confirmed with CT scanning). Venous thrombosis at the insertion site occurred in 8.7% of patients.[10]

The TGF demonstrates a tendency for crossing 1 or more pairs of filter struts. Greenfield et al reported this occurrence in 8.7% of 181 cases in 1991[11] ; the authors reported a 10% rate of filter tilt in 1994.[12] Sweeney et al reported filter tilting in 23% of patients.[13]

In vitro and clinical studies by Greenfield et al showed that clot-trapping ability and the rate of recurrent PE appeared not to be affected adversely by filter tilt or leg distribution. Other authors suggest that the jugular approach resulted in less filter tilt than the femoral approach. The TGF provides the same degree of clot-trapping ability as the stainless steel 24F Greenfield filter. The overall recurrent PE rate was reported to be 3.2-5%, which is comparable to the rates seen with other filter designs.[14] The IVC patency rate after TGF insertion was 97.8% in a long-term follow-up study of 373 patients by Greenfield et al; insertion-site thrombosis was seen in 2% of patients.[14]

Percutaneous stainless steel Greenfield filter

In the course of the continual improvement of the Greenfield filter design, a 12F stainless steel over-the-wire Greenfield filter was developed to provide a lower-profile introducing system.[11] The hook design was improved to prevent filter migration. The guidewire is used during deployment of the percutaneous steel Greenfield filter (PSGF) to provide a centering effect during filter release. Greenfield et al suspected that elimination of the guidewire may have been responsible for the high incidence of tilt of the TGF. The US Food and Drug Administration (FDA) approved the PSGF in 1995.

The PSGF, also called the stainless steel Greenfield filter with alternating hooks (SGF-AH), is manufactured from the same 316L-grade stainless steel as the original 24F SGF, and it retains the conical shape with 6 zigzag radiating legs.[15] The legs are welded to the apical bead, now at an angle of 0° as opposed to 17.5°. An opening in the center of the apical bead allows passage of a guidewire. The legs are 4.9 cm long (vs 4.7 cm for the TGF and 4.6 cm for the standard Greenfield filter). The base diameter is 3.2 cm (vs 3.8 cm for the TGF and 3.0 cm for the standard Greenfield filter). The system is preloaded differently for jugular and femoral approaches through a 12F carrier system.

Fluoroscopic control during withdrawal of the guidewire was recommended, because the guidewire can get caught between the legs near the apex or the hook. The not infrequent entrapment of guidewires was believed to result from the more crowded legs toward the apex with this new design. In 1998, Johnson et al reported that 21 (55%) of 38 patients had filter tilting of more than 15°, as evaluated with postplacement cavography and CT scanning.[16] The authors found a significant difference between filters placed via the jugular approach (12%) and filters placed via the femoral approach (51%).[16] Kinney et al also found that the right jugular approach resulted in the least filter tilt in their experience with 104 PSGF placements.[17]

In 2000, Greenfield et al reported the results of 600 PSGF placements in 599 patients.[18] Among patients receiving the PSGF with the modified hook, they found filter tilt in only 0.4%. Filter tilt was present in 5% of patients receiving the PSGF with standard hooks. Filter migration of more than 20 mm was seen in as many as 27% of PSGF placements with the standard hook but in less than 1% of the PSGF placements with alternating hooks.[18] Recurrent PE was reported to occur in 2% and 2.6% by Johnson et al and Greenfield et al, respectively. Greenfield et al reported a caval patency rate of 98% and a 4.3% rate of insertion thrombosis.

Bird's Nest filter

The Bird's Nest filter (BNF) was developed in 1982. This filter is constructed of a network of 4 biocompatible stainless steel wires (see the following image). Each wire is 25 cm long and 0.18 mm in diameter. The wires are preshaped with many nonmatching bends of a short radius and they are fixed at each end to V-shaped struts, the 2 legs of which are connected at an acute angle. A hook with a small loop stop minimizes the risk of IVC perforation at the end of each strut. The original designs called for 0.25-mm struts, but the later design (introduced in 1986) used 0.46-mm struts. This change increased the size of the loaded filter from an 8F to a 12F catheter system. The length of the introducer is 75 cm for the jugular approach and 40 cm for the femoral approach.

Bird's nest filter. Bird's nest filter.

When a Bird's Nest filter is deployed, one of the V-shaped paired struts is pushed gently to engage to the IVC wall. Originally, it was recommended that the catheter be withdrawn by 1-3 cm over the pusher wire after the hooks were fixed to the IVC; later, it was recommended that 2 or 3 twists of 360° be applied. The purpose of the twists is to prevent or reduce the chance of wire prolapse. The second pair of struts is then pushed into the IVC so that the junctions overlap by 1-2 cm. The handle of the pusher wire is turned counterclockwise 10-15 times to free it from the struts.

The Bird's Nest filter has a unique design. The final shape of the deployed filter varies owing to the flexibility of the long 0.18-mm wires. The maximum span of the V-shaped struts is 60 mm. Therefore, the Bird's Nest filter is suitable for placement in a megacava with a diameter greater than 28 mm,[19] and the FDA approved it for this use. Other filters are associated with a risk of migration if the IVC diameter is greater than 28 mm. The Bird's Nest filter has been placed successfully in patients with an IVC diameter as large as 42 mm, without filter migration.

In an experimental model, Korbin et al suggested that placement of a single Bird's Nest filter was preferred to bi-iliac placement of filters in cases of megacava.[20] Other investigators also demonstrated the clot-trapping efficacy of the BNF in various in vitro studies.[19, 21, 22, 23, 24]

Wire prolapse beyond the struts has been reported, but it does not affect the clot-trapping ability of the Bird's Nest filter. CT scan studies performed in patients with Bird's Nest filters demonstrated penetration of the caval wall in 83%. All of the patients were asymptomatic in the series reported by Wojtowycz et al.[25] The authors also reported that 8 patients (3%) in the series of 267 Bird's Nest filter insertions had thromboses at the access site.

Roehm et al reported a rate of 2.9% for IVC thrombosis in their series of 568 patients.[26] IVC thrombosis was reported to occur in as many as 17% of patients by Ferris et al,[27] in 7% by Thomas et al,[28] and in 4.7% by Nicholson et al.[29] Wojtowycz et al reported rates of 1% confirmed and 3% suspected recurrent PE, which is similar to the 2.7% rate reported by Roehm et al. Athanasoulis et al reported a 7% recurrent PE rate, 90% of which were fatal, in a series of 255 patients.[5]

Vena Tech LGM filter

The Vena Tech LGM filter was first developed and released in France by LG Medical; FDA approval was obtained in 1989. The filter is stamped and point-welded and manufactured from a biocompatible, nonferromagnetic metal alloy called Phynox (cobalt, chromium, iron, nickel, molybdenum). The filter is conical with 6 radiating, flat but slightly curved legs joining a central hub at the apex. At the base of the filter, each leg is attached to a flat side rail that contacts the wall of the IVC and thus acts as a vertical stabilizing strut.

The shape of each leg–side rail assembly resembles a skewed V or a check mark. A small spike is stamped approximately 13 mm from the cephalad end of each side rail, which provides longitudinal stability and prevents filter migration. In the original design, the side rails were 34 mm in length and the legs were 45 mm in length. In the latest design (introduced in 1991), the legs are 35.5 mm in length. The open filter is 38 mm in height. The base diameter is 30 mm and is suitable for deployment in IVCs as large as 28 mm in diameter.[30, 31] However, Millward et al reported successful insertion of a Vena Tech-LGM filter without significant migration in 4 patients with IVCs greater than 28 mm in diameter.[32] The filter is delivered through a 12F introducer sheath from either a jugular or femoral approach.

Incomplete opening of the Vena Tech LGM filter has been shown to affect the clot-trapping ability of the filter. In 1991, Reed et al reported an incomplete opening of the filter of as much as 41% when it was placed by the jugular approach, and they recommended that the filter not be inserted via this approach.[33] Murphy et al reported incomplete opening of the filter in 6% of filter insertions.[34] Rapid deployment by brisk retraction of the insertion cannula has been recommended to avoid incomplete opening.

A rate of as much as 6% of recurrent PE was reported in a number of long-term follow-up studies after insertion of the Vena Tech LGM filter. Of these patients, fatal PE occurred in 25-50%. The incidence of thrombosis at the insertion site was reported to be 8-23%. Murphy et al demonstrated a rate of 37% for filter thrombosis, with cephalad extension of thrombus seen in 20%.[34] Crochet et al reported a progressive decrease in IVC patency, with a rate of 66.8% at 9 years into the follow-up period.[35] Filter placement resulting from failed anticoagulation therapy for PE appeared to be the main predictor of IVC occlusion. Earlier, the Crochet et al reported IVC patency rates of 92% after 2 years, 80% after 4 years, and 70% after 6 years of follow-up monitoring.[36] In the same study, the authors reported an 18% rate of filter migration.

Vena Tech LP filter

The new Vena Tech filter design was approved by the FDA in 2001. Similar to the LGM, the Vena Tech LP filter is made from Phynox and, therefore, is MRI-compatible. This filter has a conical design with 8 wires for clot trapping connected to 4 sets of inverted V-shaped struts with anchoring barbs. A single design is packaged for both femoral and jugular approaches, with color-coded pushers and matching locking mechanisms for the different approaches. A newer 96-cm introducer kit for an antecubital approach is available. The introducer system is by a 9F outer diameter (OD) (7F inner diameter [ID]) sheath. When deployed, the height is 43 mm. The Vena Tech LP filter was approved by the FDA for use in patients with IVCs of diameters up to 28 mm; in Europe, it is used in patients with IVCs of up to 35 mm in diameter.

In a 2008 report on a multicenter experience in 106 patients, Le Blanche et al found preexisiting partial caval thrombosis in 5 patients and IVC occlusion in 2 patients during the first follow-up period, but no new events during the subsequent follow-up.[37] Filter tilt greater than 15° was seen in 2 patients at the 30-day follow-up. During the 3 follow-up periods, no recurrent or new PE, filter fracture, or migration was seen.[37]

Simon nitinol filter

Nitinol is an alloy consisting of 53% nickel, 45% titanium, and 2% cobalt that was developed in the 1960s at the US Naval Ordnance Laboratory. It is pliable at room temperature but has thermal memory and reverts to its preformed shape at body temperature. Nitinol resists corrosion, is nonferromagnetic, and is stronger than steel.

Simon et al performed experimental studies with this material in the 1970s, and ultimately, they developed a filter that was approved by the FDA in 1990.[38] The filter is 3.8 cm long and has 2 sections. The filter mesh, which is directed toward the heart, consists of 7 overlapping loops of 0.015-in–diameter (0.038-cm–diameter) nitinol wires fused together at 2 points. This part has an outer diameter of 28 mm. The anchoring basal part of the filter is formed by 6 diverging limbs in a radial array. Each has a small hook at the end that engages the caval wall for a depth of 1 mm (see the image below).

Simon nitinol filter. Simon nitinol filter.

The deployed filter can adapt to an IVC with a diameter as large as 28 mm. The legs also function as a coarse filter; the dome forms a fine filter and is designed to trap emboli larger than 5 mm, as well as a significant portion of smaller emboli. In vitro studies have demonstrated that the Simon nitinol filter traps more of the smaller clots than the other filters. Compared with the Greenfield filter, clot-trapping capability is not affected by filter tilt.

The Simon nitinol filter can be placed via the femoral, jugular, and antecubital[39] approaches. The delivery system for the jugular kit is longer, and the orientation of the filter is reversed in comparison with that in the femoral kit. Initially, a cold saline infusion through the storage tube via the side port was recommended to soften the filter. Later, modification of this filter eliminated the need for cold saline infusion. As a result of the flexibility of this filter at low temperature, placement through the right external jugular vein and right antecubital vein approaches have been successful using the internal jugular kit. Upon deployment, shortening of approximately 1-2 cm occurs as the dome of the filter is formed. In patients with small IVCs, incomplete formation of the dome can be seen after deployment; this does not alter filter function.

In a long-term study of 114 consecutive Simon nitinol filter insertions, Poletti et al reported a rate of PE recurrence of 4.4%.[40] Comparing different filters used during the past 26 years, Athanasoulis et al reported that the Simon nitinol filter had the lowest rate for recurrent PE (3%) and fatal recurrent PE (2%).[5] Poletti et al also reported a 3.5% rate of IVC thrombosis and a 3.5% rate of thrombosis at the femoral vein access site.[40] Aswad et al reported a higher incidence of access site and IVC thrombosis, but the rate was not significantly different from rates compared with the other types of filters studied.[41]

In a series of 103 patients, Simon et al found 5 patients with nonoccluding thrombus in the IVC, as well as 3 patients each with asymptomatic and symptomatic IVC occlusion.[42] Using MRI, Grassi et al raised the possibility of malignancy contributing to IVC thrombosis in 50% of patients with underlying malignancy undergoing Simon nitinol filter placement.[43] Filter migration or IVC perforation was not seen by Simon et al in their series of 103 patients.[42] In addition, Poletti et al found no filter migration in 38 patients in a study with a mean follow-up period of 32.2 months.[40] Strut fracture, axial deviation of the filter, and IVC perforation have been reported with the Simon nitinol filter; however, none of these complications appears to affect filter function adversely.

TrapEase filter

The TrapEase filter was approved by the FDA in the summer of 2000. The filter is laser cut from a single tube of nitinol material without welding. It has a symmetric, trapezoidal, double-basket design providing 2 levels of clot trapping, as well as self-centering characteristics. The 2 baskets are connected through 6 side struts, each with proximal and distal hooks to provide filter anchoring. The TrapEase filter is 65 mm in length before deployment and 50 mm in length when expanded maximally to a diameter of 35 mm. The design allows placement in IVCs with diameters ranging from 18 to 30 mm.[22] (See the following image.) If appropriate, the catheter can be removed endoscopically.[44]

TrapEase filter. TrapEase filter.

The unique symmetric design of the TrapEase filter allows placement from either the jugular or femoral approach by using the same 55-cm package. A separate 90-cm length package is suitable for the antecubital approach. The filter is deployed by a pusher through an 8F OD (6F ID) introducer set. Before the 90-cm kit was available, Davidson and Grassi reported 5 cases of successful antecubital placement,[45] and Stone et al reported successful placement on 39 patients via the subclavian veins.[46]

In 2001, Rousseau et al reported a 95% technical success rate from their clinical experience with the TrapEase filter in 65 patients at 12 centers in Europe and Canada; 3 filters were placed not at the intended site.[47] No filter migration, insertion site thrombosis, filter fracture, or IVC perforation was observed. At 6-months follow-up, no recurrent PE was identified. Two cases of filter thrombosis were reported.[47] Other authors reported similar results. Of 751 TrapEase filters placed, Kalva et al reported 2 cases (0.2%) of near-total caval occlusion and 15 cases (6.8%) of CT scan–confirmed PE in 219 patients.[48]

SafeFlo filter

The SafeFlo filter has a unique design; it is made of nitinol wires, and the anchoring mechanism is a double-ring platform. These 2 rings are oversized to exert radial force on the IVC and are slightly angled toward each other to provide additional bidirectional stability. This is connected to the filter element that is designed to be 3 mm smaller than the IVC. It is shaped from nitinol wires into an outer supporting ring and a 5-leafed inner filtration configuration.

The SafeFlo filter is offered in 3 sizes to fit IVC diameters from 15 to 27 mm. Deployment is through a 7F sheath via either a jugular or femoral approach.

Retrieval can be performed through the use of the VascuGrab retrieval system; however, the VascuGrab system is not yet approved by the FDA and not available in the United States.

In vitro results were previously published by Bruckheimer et al.[49] A mutlicenter clinical study has been performed, and the clinical results are pending.

Retrievable/optional IVC filters

Over the past decade, several new filters have been approved as retrievable filters. These devices are sometimes referred to as optional filters, because they can be retrieved after a period following placement or can be left in the patient as a permanent device.[50, 51, 52, 53, 54, 55, 56, 57, 58, 59]

Retrievable/optional IVC filters include the following:

  • Günther Tulip filter
  • Recovery filter
  • Recovery G2/G2 and G2 X filters
  • OptEase filter
  • Celect filter
  • ALN filter
  • Option filter

The SafeFlo filter is available in Europe as an optional filter, but it is currently approved only as a permanent filter in the United States.

There are also temporary filters that require removal, such as the Tempo filter II. Convertible filters are optional filters that the filtration property can be altered with a second procedure. Such an example is the VenaTech Convertible filter, which becomes a stent after conversion. These are currently not available in the United States.

Some patients need a filter to protect from PE for only a short period. Decousus et al suggested that IVC filters provide only short-term protection.[60] Significant reduction of the PE rate was reported at day 12, but no difference was reported at 2 years when compared with anticoagulation therapy. After over 8 years of follow-up, the incidence of postthrombotic syndrome and of DVT were higher in the filter group, and the PE rate was lower.[60] In the same follow-up study by the Prevention du Risque d’Embolie Pulmonaire par Interruption Cave, a cumulative incidence of symptomatic IVC thrombosis of 13% after 8 years was noted[60] ; by contrast, Crochet et al reported a higher rate of 33% caval thrombosis in a 9-year follow-up of the Vena Tech LGM filter.[35]

The Society of Interventional Radiology guidelines for the use of retrievable filters recommends that "the decision to use an optional filter rather than a permanent filter should be based on the anticipated required duration of protection against clinically significant PE and/or risk of pharmacologic therapy."[54]

When the indication for caval filtration is no longer present in a patient or the risk of PE is acceptably low because of a change in clinical status or because prophylactic therapy was successful, these patients can be considered for filter retrieval. Patients should be fully evaluated by clinical status, laboratory findings, and imaging before the attempted retrieval.

When there is a need for continuous caval filtration, optional filters can be left in situ as a permanent device. In some cases, these filters are not removed or their retrieval may be deferred for various reasons, such as poor clinical status, short lifespan, significant amount of clot trapped within the filter, IVC thrombosis, or technical difficulties that prevent safe retrieval.

In younger patients who have a longer life expectancy, use of a retrievable filter is a more valid option for physicians and patients than the use of a permanent device. These optional filters have gained popularity over the years. More and more of these devices are being left as permanent filters.[61] Although the percentage of successful retrieval varies, some series quote retrieval rates of up to 100% technical success for the selected patients.

Günther Tulip filter

The Günther Tulip filter was approved by the FDA in 2003.[62, 63, 64, 65, 66, 67, 68, 69, 70, 71, 72, 73] This filter is made from nonferromagnetic Conichrome, and it is conical in design with a hook at the apex for retrieval. It has 4 legs of 44 mm with barbed hooks for anchoring; 4 additional wires wrap around the primary legs to form a tulip-petal conical structure with a total of 12 wires for clot trapping. The maximum diameter is 30 mm, and its maximum height is 50 mm. The introducer system is of 8.5F ID, and different packages are available for femoral or jugular approaches. The retrieval system is of 7F ID (11F OD for the sheath) and is employed via a jugular approach only.

The direction of the hook is a factor that influences the rate of success during retrieval of the Günther Tulip filter. Millward et al recommend orientating the hook toward the lumen of the IVC, especially if filter tilting is anticipated. These authors also advocate anticoagulation for 24-48 hours before retrieval.

A venacavogram must be performed before retrieval to assess the amount of trapped clot. More than 25% of clot trapped in the cone is a contraindication to retrieval. The filter is also not to be retrieved in cases of continued requirement of a filter or if it is technically not possible to be retrieved because of trapped thrombus, adherence to the IVC wall, or filter tilt. The reported total caval occlusion rates are 0-6%.

The recommended retrieval time is in the first 12 days, but various authors have prolonged the time to retrieval with or without repositioning the filter for up to 3-4 weeks, or even longer.[74]

In 2009, Smouse et al reported on a multicenter experience of 554 patients. Filter retrieval was successful in 248 (90.2%) of 275 patients.[74] The mean indwell time was 58.9 days (range, 3-494 d). Their analysis showed a probability of successful retrieval greater than 94% at 12 weeks and greater than 67% at 26 weeks.[74] Excessive filter tilt and tissue in-growth are main causes of failed retrieval.

Recovery nitinol filter

The Recovery filter was approved by the FDA in 2003, but it is no longer available commercially and has been replaced by the G2/G2 X filters (discussed further below).

The Recovery filter is made from nitinol and is MRI-compatible. Similar to the Simon nitinol filter, it has 2 levels of filtration using 6 radiating arms and legs, each of 0.13-in nitinol wires. The legs have hooks for anchoring. The maximum diameter of the extended filter legs is 32 mm; it is approved for use in IVCs of up to 28 mm in diameter. The deployed height is 40 mm. The delivery system is of 7F ID (9F OD) through the femoral approach.

Retrieval is via a specific retrieval cone supplied as a separate kit. The system is introduced via the jugular route with a 12F sheath. The 20-mm cone has 9 steel struts within a urethane membrane with a base diameter of 15 mm when opened.

Asch reported successful retrieval of the Recovery filter without complications in 24 patients with a mean implantation period of 53 days (range, 5-134 d).[75] One migration of 4 cm occurred with a large, trapped thrombus. Grande et al reported recurrent PE in 2.8% of 106 Recovery filters placed[76] ; 14 filters were retrieved successfully from this group.

Kalva et al report their experience in 96 patients, in which retrieval was successful in 9 of 11 attempted patients.[77] The average indwell time was 117 days (range, 24-426d). On abdominal CT scan in 40 of the 96 patients, 12 (30%) showed clumping of the filter legs, and 11 (27.5%) showed penetration of the IVC walls.

In a series of 14 patients, Hull and Robertson reported arm perforation in 5 of 9 (56%) and leg perforation in 1 of 9 (11%) on early CT imaging (mean, 30 d; range, 0-126 d) and in all patients during final imaging (mean, 899 d; range, 119-1218 d).[78] They reported limb fracture and migration in 3 of their 14 patients (21%) and successfully performed late-stage retrieval on all the filters.[78] Cantwell et al reported 6 filter fractures in 128 patients.[79]

Recovery G2/G2 & G2 X filters

The Recovery G2 (or simply G2) and G2 X filters are modifications of the original Recovery filter.[80] These were approved by the FDA in 2008 and 2009, respectively.

The redesigned G2 filter has enhanced fracture resistance, better centering, improved fixation hooks, and longer legs with a wider span to reduce migration for IVCs up to 28 mm diameter. It is packaged separately for femoral and jugular approaches. The femoral kit has a 7F ID introducer sheath, whereas the jugular kit has a 10F ID set. The deployed height is 42 mm.

The retrieval set is the same as for the original Recovery filter. The G2 X filter has a hook at the apex, and is therefore retrievable by both the snare technique or with the cone retrieval kit.

Kim et al reported 5 new, symptomatic PE (6.5%) in 77 G2 filters placed.[50] In 2008, of 59 patients evaluated, Oliva et al reported 100% success for retrieval in 51 patients, with a mean dwell time of 53.4 days (range, 7-242 d).[81] Of these, 8 of 59 cases showed large trapped thrombi during cavography. Oliva et al reported 9 of 51 (18%) patients with IVC wall penetration and 6 of 51 (12%) patients with filter tilt greater than 15o.[81] There were no fractures or cranial migrations, but 2 caudal migrations occurred.

Charles et al reported filter tilt in 5 (18.5%) of their 27 patients.[82] In 2009, Binkert at al reported 15% filter tilts, 12% caudal migrations, and 1.2% filter fractures in 85 patients on a 6-month follow up.[83] Of the 61 patients referred for retrieval, 16 (26%) showed presumed penetration cavography, and 14 of 16 had successful retrieval.[83]

In the same year, Cantwell et al reported no filter fractures in their series of patients that compared the Recovery and G2 filters.[79] They further concluded that the G2 had less filter tilt but a higher rate of caudal migration than the Recovery filter.[79] Lynch and Kekulawela reported 6 (3.4%) filter fractures in 174 G2 filters placed, all of which were implanted for over 180 days.[84] They observed 76 of 174 (44%) cases of presumed caval penetration during removal.[84]

There have also been 2 reports of intraventricular migration.[85, 86]

OptEase filter

The OptEase filter is similar in design to the permanent TrapEase filter. It is cut from a single tube of nitinol, with a 6-sided, conical shape and dual-level clot-trapping capability. Although the OptEase filter can be deployed via either a femoral or jugular approach, the provision of a hook on the caudal end and unidirectional anchoring barbs to prevent cephalad migration allow retrievability from the femoral approach, which is unique to this design.

In 2002, this filter was approved by the FDA for retrieval times of up to 23 days. The introducer system has a profile similar to the TrapEase filter; the retrieval system is 10F.

Kim et al reported 11 cases (4.2%) of new PE and 1 case (0.4%) of IVC thrombosis in 260 OptEase filters placed.[87] Keller et al reported on removal of 58 of 80 OptEase filters placed,[88] in which 5 patients required additional access during retrieval, because the hooklet could not be engaged. The IVC thrombosis rate in their series was 3%.[88] No IVC thrombosis was seen in 27 patients reported by Oliva et al,[89] or in 94 patients reported by Rosenthal et al.[90] However, Corriere et al reported 3 (12.5%) cases of IVC thrombosis in 24 patients with retrievable filters; all 3 were OptEase filters.[91]

Celect filter

The Celect filter is a newer retrievable filter approved by the FDA in 2008. It is based on the Günther Tulip filter. The new design is made of the same material, Conichrome. The Celect filter has 4 anchoring struts for fixation and 8 independent secondary struts to improve self-centering and clot trapping.

The deployment kit is similar to that for the Günther Tulip filter, and it is available for both the femoral and jugular approaches. The same Günther Tulip filter retrieval set is employed for retrieval from the jugular approach.

Initial studies suggest that the retrieval interval of the Celect filter can be extended, with higher rates of successful retrieval.[92, 93]

In 2009, in a group of patients with Günther Tulip and Celect filters placed by intravascular ultrasonography (IVUS), Rosenthal et al reported filter tilt in 15 of 50 (30%) for the Günther Tulip filter and 6 of 65 (9.2%) for the Celect filter at cavography before retrieval.[94] Sangwaiya et al reported successful retrieval in 14 of 15 patients (93.3%) in a group of 73 placements, with a mean indwell time of 84 days (range, 5-381 d).[95] Two (2.8%) patients developed new PE after filter placement, similar in incidence to previous reports, and penetration of filter legs or arms was seen in 4 (22.2%). Fracture and migration of filter component, filter thrombosis, and iliac vein thrombosis were seen in 1 patient each (5.5%).[95]

In the same year, in a multicenter experience in 95 patients with Celect filter placement, Lyon et al reported 56 (96.6%) of the filters were successfully retrieved out of 58 patients with a mean indwell time of 179 days (range, 5-466 d).[96] They observed filter tilt greater than 15° in 5 (8.6%) patients at deployment and 2 (3.4%) patients at retrieval. Strut penetration was observed in 21 (36%) patients but no IVC perforation; 20 of these were retrieved successfully.[96] No significant filter migration, filter fracture, or IVC occlusion was observed.

ALN filter

The ALN filter has been available in Europe since 1997. It was approved by the FDA in 2008.

This filter is a conical filter made from 9 variable-length wires of 316L stainless steel. The upper level has 6 arms with hooks for anchoring and 3 longer legs for centering. The introducer sheaths are 7F, and the filters are preloaded in a holder. The filter kits are packaged separately for deployment for femoral, jugular, and brachial approaches.

Retrieval is via the jugular approach. The introducer set is 9F OD. The retrieval kit is a forceps design, cone-retrieval system with 8 stainless wires without any plastic cone.

In 2007, in a series of 220 patients with ALN filter placement with an 18-month follow-up, Mismetti et al reported filter tilt was seen in 12 of 212 (5.7%) during placement.[97] Fatal PE occurred in 5 (2.3%) within 24 hours to 11 days following placement. Filter retrieval was successful in 55 patients, with a mean indwell time of 51 days (range, 6-352 d); 4 of these patients required a second attempt because of filter thrombosis.[97]

In 2008, Pellerin et al reported successful retrieval was achieved in 122 of 123 patients (99%), with a mean indwell time of 93 days (range, 6-722 d).[57] Severe tilting was seen in 6 cases (5%), including 1 failure in retrieval. There was no filter migration, fracture, or IVC penetration.[57]

Option filter

The Option filter was approved by the in FDA in June 2009.

This is a laser-cut nitinol device designed for IVC diameters up to 32 mm. A somewhat unique feature is over-the-wire placement. The Option filter is a conical filter featuring 6 legs with hooks for anchoring, with a retrieval hook at the apex. The introducer set is 6F OD (5F ID), and the filter cartridge is color-coded for a femoral/jugular approach.

Clinical results are pending.

Previous
Next

Indications

Deep venous thrombosis (DVT) and pulmonary embolism (PE) represent two points on the continuum of a single disease process. The reported annual incidence of DVT in the United States includes a wide range (250,000-20,000,000 cases of lower-extremity DTV involving treatment or hospitalization), and the disease constitutes a major healthcare burden.

Most commonly, DVT develops within the deep veins of the lower extremities, but it can also involve, or arise solely from, the veins of the pelvis or the upper extremities. The most feared complication of DVT is PE. PE frequently results in hospitalization, and it is associated with high morbidity and mortality rates. Currently, PE is the third leading acute cardiovascular cause of death in the United States. Death is associated with serious underlying disease in approximately 50% of patients with PE.

Systemic anticoagulation with intravenous heparin followed by oral warfarin remains the mainstay of treatment for DVT and for prevention of PE. However, studies have shown that as many as 33% of patients develop a second PE while receiving adequate anticoagulation therapy.

In addition, anticoagulation therapy is associated with hemorrhage, which precludes its use in certain groups of high-risk patients, including patients at high risk for falling, hemorrhagic stroke, central nervous system (CNS) metastasis, or bleeding diathesis. Although heparin can be used during pregnancy, warfarin crosses the placenta and has been shown to have significant adverse effects on the developing fetus.

The concept of interrupting the flow in the inferior vena cava (IVC) to prevent PE originated in the 1930s-1940s with the performance of ligation of the common femoral vein and superficial femoral vein. These methods were used in parallel with anticoagulation methods when heparin and warfarin became available in 1935 and 1948, respectively. Phlebotomy with clot removal was also performed if thrombus was present at the common femoral vein level. A high incidence of limb edema was associated with femoral vein ligation.

Later, ligation of the IVC was performed, initially in patients harboring thrombi above the superficial femoral veins. Eventually, it replaced lower-level venous ligation. The operative mortality rate of 2-15% was not significantly different from that for femoral vein ligation, but a lower rate for recurrent PE was reported. The optimal level of ligation was located immediately below the renal veins to prevent thrombosis caused by venous stasis between the interrupted IVC and the renal veins. However, 10-16% of these patients had immediate lower-extremity swelling.

In the 1960s, various methods of partially interrupting the flow in the IVC were developed to lessen the effect of venous stasis. The procedures provided emboli trapping with preservation of blood flow through the lumen. The procedures included suture plication and the use of caval clips (Moretz clip, Miles clip, Adams-DeWeese clip). The rates for operative mortality and recurrent PE were similar to those for caval ligation, yet the rate for limb edema was reduced. Despite the fact that these procedures partially interrupted flow in the IVC, caval occlusion rates of 30-40% were reported.

In 1967, the Mobin-Uddin umbrella filter (see Design Features) was developed as a replacement for surgical ligation, caval plication, and caval clips, as well as partially to interrupt the flow in the IVC and to prevent PE. Since then, several filters have been introduced. Currently, filters represent the standard of care when partial interruption of IVC flow is indicated to prevent the occurrence of PE.

DVT or PE in a patient with contraindication to anticoagulation therapy

Although the indications for IVC filter placement have expanded over the years, DVT or PE in a patient for whom anticoagulation therapy is contraindicated remains the most frequent indication, accounting for 38-77% of patients undergoing IVC filter placement.

Contraindications to anticoagulation therapy include the following:

  • Hemorrhagic stroke
  • Recent neurosurgical procedures or other major surgery
  • Major or multiple trauma
  • Active internal bleeding (eg, upper or lower gastrointestinal bleeding, hematuria, hemobilia)
  • Intracranial neoplasm
  • Bleeding diathesis (eg, secondary thrombocytopenia, idiopathic thrombocytopenic purpura, hemophilia)
  • Pregnancy
  • Unsteady gait or tendency to fall (as seen in patients with previous stroke, Parkinson disease)
  • Poor patient compliance with medications

DVT or PE in a patient with a complication of anticoagulation therapy

DVT or PE in a patient who has a complication of anticoagulation therapy accounts for 6-18% of IVC filter placements.

Anticoagulation therapy is not without potential complications; it is associated with a mortality of 5-12%, and therapy may have to be discontinued in as many as 18% of patients because of hemorrhage. Anticoagulation with heparin is also associated with immune-mediated heparin-induced thrombocytopenia (HIT) in 1-3% of treated patients. Approximately 25% of patients with HIT develop new venous and/or arterial thrombotic events, including PE.

Failure of anticoagulation therapy

Patients who have new-onset PE or DVT while they are receiving anticoagulation therapy are candidates for IVC filter placement. Among patients with PE, 3-33% have a second embolic event during adequate intravenous heparin therapy or oral warfarin anticoagulation treatment.

Free-floating iliofemoral or caval thrombus

Free-floating iliofemoral or IVC thrombus has been associated with a 27-60% rate of PE. Therefore, it is considered an indication for IVC filter placement. However, results have not been consistent. For example, Pacouret et al did not find a higher risk for PE in patients with free-floating proximal DVT.[98]

PE prophylaxis

IVC filter placement has been advocated as a means of preventing PE in patients at high risk for thromboembolic events. Traditionally, such patients have included the following populations:

  • Patients with DVT who are about to undergo surgery (lower-extremity orthopedic surgery, major abdominal surgery, neurosurgery)
  • Patients with chronic pulmonary hypertension and a marginal cardiopulmonary reserve
  • Patients with cancer
  • Trauma patients, including those with (1) severe head injury with prolonged ventilator dependence, (2) major abdominal or pelvic penetrating venous injury, (3) spinal cord injury with or without paralysis, (4) severe head injury with multiple lower-extremity fractures, or (5) pelvic fracture with or without lower-extremity fractures

In 1981, Moore et al reported 21 hemorrhagic events in 32 patients with malignancy who underwent anticoagulation therapy to treat thromboembolic disease.[99] Of these patients, 8 had hemorrhage resulting in cessation of therapy or death. This observation led to the conclusion that IVC filter placement is a safer means for PE prophylaxis.[99] However, this indication was disputed in 1998 by the American College of Chest Physicians Consensus Committee on Pulmonary Embolism[100] ; its report stated that the routine use of IVC filters is not recommended in patients with cancer and DVT or PE.[101]

Patients with severe trauma are prone to develop DVT and PE. In many of these patients, anticoagulation therapy is contraindicated because of the risk of hemorrhage. Other conservative methods, including compression devices and foot pumps, may not adequately prevent DVT. Although some studies have shown that the prophylactic placement of IVC filters prevents fatal PE in patients with trauma, other studies have not shown any significant reduction in the rate of PE. Additional studies of cost-effectiveness or risk-benefit considerations do not support prophylactic filter placement in patients with trauma.[102]

Contraindications

IVC filter placement has been shown to be safe, with only a few relative contraindications in some patients, such as the following:

  • Those receiving therapeutic anticoagulants
  • Those with thrombus between the venous access site and expected deployment site
  • Patients expected to undergo magnetic resonance imaging (MRI) after filter placement

Many of these contraindications have been addressed with changes in technique and the type of filter used. With smaller introduction and deployment systems and with the availability of ultrasonographic (US) guidance during venous puncture, the reversal of anticoagulation often is not necessary.

The presence of thrombus between the puncture and deployment sites requires the selection of a filter that allows an alternate approach (ie, brachial or jugular vein). The titanium Greenfield, Vena Tech-LGM, Vena Tech LP, Simon nitinol, TrapEase, OptEase, Günther Tulip, and Recovery filters are nonferromagnetic; therefore, for patients with these filters, subsequent exposure to the magnetic field during MRI is possible (see Design Features).

Previous
Next

Clinical Implementation

This section discusses normal and variations of inferior vena cava (IVC) anatomy, preprocedural evaluations in assessing deep vein thrombosis (DVT) and pulmonary embolism (PE), and assessing the IVC for clinical implementation of IVC filters. See also Inferior Vena Cava Vascular Filter Placement.

Normal anatomy, IVC

The IVC is the largest venous structure in the body. It drains the venous return from the lower extremities, pelvis, and abdomen into the right atrium. The IVC originates at the L4-5 vertebral level and is formed by the confluence of the left and right common iliac veins. At this level, the IVC is posterior to the origin of the right common iliac artery. As it ascends to the right of the aorta and anterior to the spine retroperitoneally, the IVC receives blood from the lower lumbar veins and ascending lumbar vein.

The right gonadal vein and suprarenal veins empty directly into the IVC below and above the right renal vein, respectively. The counterparts of these veins on the left side drain into the IVC via the left renal vein. The renal veins join the IVC at vertebral level L1-2. The right renal vein has a shorter course and is typically situated more caudally. The left renal vein is longer and typically crosses in front of the aorta. At the level of the renal veins, the IVC lies posterior to the head of the pancreas. Then, it ascends in a groove posterior to the liver. The IVC receives the left, middle, and right hepatic veins, and often a separate branch from the caudate lobe, before it passes behind the right crus of the diaphragm to enter the right atrium (see the image below).

Anteroposterior image from an inferior venacavogra Anteroposterior image from an inferior venacavographic examination demonstrates an inferior vena cava of normal diameter without thrombus. In addition, the inflow from the renal veins can be seen at the level of the L1 vertebral body.

Anatomic variations, IVC

Anatomic variations involving the IVC are fairly common in the general population, reflecting variation in the involution and persistence of the cardinal veins. These conditions may be important, because they alter the location of filter deployment in 3-15% of patients. A few of the more common conditions, especially those pertinent to IVC filter placement, are described below.

Duplication of the IVC

Duplication of the IVC is observed in 0.2-3% of individuals. With this anomaly, the left IVC is usually smaller than the right, although they can be equal in size. The left IVC typically empties into the left renal vein (see the following image) and, subsequently, into the right IVC. Although the frequency of duplication is low, demonstration of the anomaly is important when determining the position for IVC filter placement. When the IVC is duplicated, filter placement in the right IVC may not be adequate to prevent PE, especially in the presence of a left lower-extremity DVT. Placement of a filter in both the left and right IVCs is often required for true PE prophylaxis in the patient with a duplicated IVC.

Anteroposterior image from an inferior venacavogra Anteroposterior image from an inferior venacavographic examination demonstrates duplication of the inferior vena cava. The left-sided inferior vena cava drains into the left renal vein.

Transposition of the IVC

Transposition of the IVC is observed in 0.2-0.5% of individuals. A left-sided IVC drains into the left renal vein and crosses to the right of the spine. It then continues cranially in the normal position. The left renal and suprarenal veins empty directly into the left IVC; the right gonadal and suprarenal veins drain into the right renal vein and, subsequently, into the normal right-sided prerenal division of the left-sided IVC.

Circumaortic and retroaortic left renal vein

The incidence of circumaortic renal veins has been reported to be as high as 8.7% (see the image below). Patients with circumaortic or multiple renal veins may have a large hilar communication, which provides an alternative conduit for emboli. In the presence of a circumaortic renal vein, a filter immediately inferior to the main left renal vein may not represent adequate prophylaxis because of the alternative conduit; therefore, filters should be positioned inferior to the circumaortic renal vein. A retroaortic left renal vein is more common but does not impact filter placement, because it does not represent a potential conduit for thrombus arising from the lower extremities.

Anteroposterior image from an inferior venacavogra Anteroposterior image from an inferior venacavographic examination demonstrates a circumaortic left renal vein.

Preprocedural evaluations, assessing DVT and PE

Candidates for IVC filter placement are typically evaluated for DVT and/or PE. Imaging methods used to document lower-extremity DVT include ultrasonography (US) and peripheral venography. PE is usually diagnosed by nuclear medicine ventilation-perfusion (V/Q) scans.

Although pulmonary arteriography remains the criterion standard for confirmation of diagnosis, an increasing number of centers are using multichannel detector-row helical (MDH) computed tomography (CT) to evaluate patients in whom PE is suspected. Alternatively, magnetic resonance angiography (MRA) of the pulmonary vasculature can also be performed.

Laboratory assessment

Laboratory assessment typically includes the following:

  • Coagulation profile (including platelet count)
  • Prothrombin time (PT)
  • Activated partial thromboplastin time (aPTT)
  • Renal function with serum blood urea nitrogen (BUN) and creatinine levels

If a filter is to be placed in a patient who is receiving anticoagulation therapy, the medications are usually discontinued before the procedure. Discontinuing heparin on call is usually adequate, although some physicians prefer to discontinue heparin for 2-4 hours before the procedure.

Gray et al described the safety of venous interventional procedures, including pulmonary angiography and filter placement, in 100 patients receiving anticoagulation.[103] Of these patients, 87 had prolonged bleeding times.[103]

Assessing the IVC

The caliber of the IVC varies in accordance with the intravascular volume, the phases of respiration, and cardiac function. The mean diameter of the IVC is 19-20 mm; approximately 3% of patients have a megacava, in which the diameter is greater than 28 mm.

Imaging and cavography have been used to visualize the IVC for its size, its configuration, and any anatomic variation. Imaging is also used to evaluate the presence and extent of bland and tumor thrombus within the IVC.

Ultrasonography

The usefulness of US sometimes can be limited by overlying bowel gas and a large body habitus. This imaging modality is used more commonly in the evaluation of the common femoral veins or the internal jugular veins to assess patency before puncture and for guidance during venous access. More recently, intravascular US (IVUS) has been used as an imaging method before and during filter placement.[104]

CT scanning and MRI

CT scanning and MRI have been used to visualize the size, configuration, and anatomic variants of the IVC.

Inferior vena cavography

Currently, cavography is the modality used most commonly to assess the IVC, usually before filter placement. Access to the IVC is typically achieved via a left or right femoral vein or via the right internal jugular vein. The right femoral vein is the most common approach, accounting for 69% in a series reported by Athanasoulis et al.[5] US may be used to guide the puncture. After access into the venous system is gained, a pigtail catheter is typically positioned near the confluence of the common iliac veins.

Alternatively, catheter placement in the left common iliac vein enables visualization of both a normal IVC and a duplicated IVC, which is a normal variation of a double IVC.

In patients at intermediate risk for an adverse reaction to the contrast material and in patients with compromised renal function, cavography can be performed with carbon dioxide.

Various methods can be used to determine the diameter of the IVC, especially when digital subtraction angiography is performed. These include placement of a ruler with metallic markers along the left side of the patient's body and the use of marker catheters or guidewires.

In most centers, a single frontal projection is used for measurement, but biplane studies can also be used during the procedure.

More recently, investigators have advocated the use of preprocedural cross-sectional imaging for a more accurate assessment of the caval diameter.

Previous
Next

Complications

Postprocedure complications may be related to jugular puncture or to the inferior vena cava (IVC) filter itself.

Complications related to jugular puncture

The following are complications associated with jugular puncture:

  • Pneumothorax (incidence very low with ultrasonographic guidance) [105]
  • Access site thrombosis (2%)
  • Persistent bleeding from insertion site
  • Pulmonary embolism (PE) (if passage was through a thrombosed vein)

Complications related to filter

The following are complications associated with IVC filters:

  • IVC trauma
  • Penetration of the caval wall by the filter legs (may rarely lead to retroperitoneal hematoma or bowel injury)
  • Filter migration (< 1%; may be caused by trapped guidewires)
  • Filter fracture (< 1%)
  • Filter infection (< 1%)
  • Caval thrombosis (5%)
  • Death (30-d mortality, < 1%)
Previous
 
Contributor Information and Disclosures
Author

Gary P Siskin, MD Professor and Chairman, Department of Radiology, Albany Medical College

Gary P Siskin, MD is a member of the following medical societies: American College of Radiology, Society of Interventional Radiology, Cardiovascular and Interventional Radiological Society of Europe, Radiological Society of North America

Disclosure: Nothing to disclose.

Coauthor(s)

Bartholomew Kwan, MBBS, FRCPC, FRCR Staff Radiologist, Department of Medical Imaging, WOHS Brampton Civic Hospital

Bartholomew Kwan, MBBS, FRCPC, FRCR is a member of the following medical societies: American Roentgen Ray Society, Society of Interventional Radiology, Cardiovascular and Interventional Radiological Society of Europe, Radiological Society of North America, Royal College of Physicians and Surgeons of Canada, Royal College of Radiologists

Disclosure: Nothing to disclose.

Chief Editor

Kyung J Cho, MD, FACR, FSIR William Martel Professor of Radiology, Interventional Radiology, University of Michigan Health System, Frankel Cardiovascular Center

Kyung J Cho, MD, FACR, FSIR is a member of the following medical societies: American College of Radiology, American Heart Association, American Medical Association, American Roentgen Ray Society, Association of University Radiologists, Radiological Society of North America

Disclosure: Nothing to disclose.

Additional Contributors

Anthony Watkinson, MD Professor of Interventional Radiology, The Peninsula Medical School; Consultant and Senior Lecturer, Department of Radiology, The Royal Devon and Exeter Hospital, UK

Anthony Watkinson, MD is a member of the following medical societies: Radiological Society of North America, Royal College of Radiologists, Royal College of Surgeons of England

Disclosure: Nothing to disclose.

References
  1. Taccone FS, Bunker N, Waldmann C, et al. A new device for the prevention of pulmonary embolism in critically ill patients: Results of the European Angel Catheter Registry. J Trauma Acute Care Surg. 2015 Sep. 79(3):456-62. [Medline].

  2. Greenfield LJ, McCurdy JR, Brown PP, Elkins RC. A new intracaval filter permitting continued flow and resolution of emboli. Surgery. 1973 Apr. 73(4):599-606. [Medline].

  3. Tadavarthy SM, Castaneda-Zuniga W, Salomonowitz E, et al. Kimray-Greenfield vena cava filter: percutaneous introduction. Radiology. 1984 May. 151(2):525-6. [Medline].

  4. Greenfield LJ, Proctor MC. Twenty-year clinical experience with the Greenfield filter. Cardiovasc Surg. 1995 Apr. 3(2):199-205. [Medline].

  5. Athanasoulis CA, Kaufman JA, Halpern EF, Waltman AC, Geller SC, Fan CM. Inferior vena caval filters: review of a 26-year single-center clinical experience. Radiology. 2000 Jul. 216(1):54-66. [Medline].

  6. Kantor A, Glanz S, Gordon DH, Sclafani SJ. Percutaneous insertion of the Kimray-Greenfield filter: incidence of femoral vein thrombosis. AJR Am J Roentgenol. 1987 Nov. 149(5):1065-6. [Medline].

  7. Proctor MC, Greenfield LJ, Cho KJ, Moursi MM, James EA. Assessment of apparent vena caval penetration by the Greenfield filter. J Endovasc Surg. 1998 Aug. 5(3):251-8. [Medline].

  8. Greenfield LJ, Cho KJ, Proctor M, et al. Results of a multicenter study of the modified hook-titanium Greenfield filter. J Vasc Surg. 1991 Sep. 14(3):253-7. [Medline].

  9. Greenfield LJ, Cho KJ, Tauscher JR. Evolution of hook design for fixation of the titanium Greenfield filter. J Vasc Surg. 1990 Sep. 12(3):345-53. [Medline].

  10. Xenos ES, Minion DJ, Sorial EE, Endean ED. Endovascular retrieval of an intraaortic greenfield vena cava filter. Vasc Endovascular Surg. 2008 Apr-May. 42(2):165-7. [Medline].

  11. Greenfield LJ, Tauscher JR, Marx MV. Evaluation of a new percutaneous stainless steel Greenfield filter by intravascular ultrasonography. Surgery. 1991 Jun. 109(6):722-9. [Medline].

  12. Greenfield LJ, Proctor MC, Cho KJ, et al. Extended evaluation of the titanium Greenfield vena caval filter. J Vasc Surg. 1994 Sep. 20(3):458-64; discussion 464-5. [Medline].

  13. Sweeney TJ, Van Aman ME. Deployment problems with the titanium Greenfield filter. J Vasc Interv Radiol. 1993 Sep-Oct. 4(5):691-4. [Medline].

  14. Greenfield LJ, Proctor MC, Cho KJ, Wakefield TW. Limb asymmetry in titanium Greenfield filters: clinically significant?. J Vasc Surg. 1997 Nov. 26(5):770-5. [Medline].

  15. Greenfield LJ, Proctor MC, Roberts KR. An improved process for development and testing of vena caval filters: the percutaneous steel Greenfield filter. Surgery. 1997 Jan. 121(1):50-7. [Medline].

  16. Johnson SP, Raiken DP, Grebe PJ, Diffin DC, Leyendecker JR. Single institution prospective evaluation of the over-the-wire Greenfield vena caval filter. J Vasc Interv Radiol. 1998 Sep-Oct. 9(5):766-73. [Medline].

  17. Kinney TB, Rose SC, Weingarten KE, Valji K, Oglevie SB, Roberts AC. IVC filter tilt and asymmetry: comparison of the over-the-wire stainless-steel and titanium Greenfield IVC filters. J Vasc Interv Radiol. 1997 Nov-Dec. 8(6):1029-37. [Medline].

  18. Greenfield LJ, Proctor MC, Michaels AJ, Taheri PA. Prophylactic vena caval filters in trauma: the rest of the story. J Vasc Surg. 2000 Sep. 32(3):490-5; discussion 496-7. [Medline].

  19. Reed RA, Teitelbaum GP, Taylor FC, Pentecost MJ, Roehm JO. Use of the Bird's Nest filter in oversized inferior venae cavae. J Vasc Interv Radiol. 1991 Nov. 2(4):447-50. [Medline].

  20. Korbin CD, Reed RA, Taylor FC, Pentecost MJ, Teitelbaum GP. Comparison of filters in an oversized vena caval phantom: intracaval placement of a bird's nest filter versus biiliac placement of Greenfield, Vena Tech-LGM, and Simon nitinol filters. J Vasc Interv Radiol. 1992 Aug. 3(3):559-64. [Medline].

  21. Teitelbaum GP, McKay RH, Katz MD. Bird's nest filter placement within an enlarged hemiazygos vein for prevention of pulmonary embolism. Cardiovasc Intervent Radiol. 1993 Mar-Apr. 16(2):119-21. [Medline].

  22. Shlansky-Goldberg R, Wing CM, LeVeen RF, Cope C. Effectiveness of a prolapsed bird's nest filter. J Vasc Interv Radiol. 1993 Jul-Aug. 4(4):505-11. [Medline].

  23. Roehm JO Jr, Thomas JW. The twist technique: a method to minimize wire prolapse during Bird's Nest filter placement. J Vasc Interv Radiol. 1995 May-Jun. 6(3):455-9. [Medline].

  24. Starok MS, Common AA. Follow-up after insertion of Bird's Nest inferior vena caval filters. Can Assoc Radiol J. 1996 Jun. 47(3):189-94. [Medline].

  25. Wojtowycz MM, Stoehr T, Crummy AB, McDermott JC, Sproat IA. The Bird's Nest inferior vena caval filter: review of a single-center experience. J Vasc Interv Radiol. 1997 Mar-Apr. 8(2):171-9. [Medline].

  26. Roehm JO Jr, Gianturco C, Barth MH, Wright KC. Percutaneous transcatheter filter for the inferior vena cava. A new device for treatment of patients with pulmonary embolism. Radiology. 1984 Jan. 150(1):255-7. [Medline].

  27. Ferris EJ. George W. Holmes Lecture. Deep venous thrombosis and pulmonary embolism: correlative evaluation and therapeutic implications. AJR Am J Roentgenol. 1992 Dec. 159(6):1149-55. [Medline].

  28. Thomas JH, Cornell KM, Siegel EL, Sparks C, Rosenthal SJ. Vena caval occlusion after bird's nest filter placement. Am J Surg. 1998 Dec. 176(6):598-600. [Medline].

  29. Nicholson AA, Ettles DF, Paddon AJ, Dyet JF. Long-term follow-up of the Bird's Nest IVC Filter. Clin Radiol. 1999 Nov. 54(11):759-64. [Medline].

  30. Millward SF, Peterson RA, Moher D, et al. LGM (Vena Tech) vena caval filter: experience at a single institution. J Vasc Interv Radiol. 1994 Mar-Apr. 5(2):351-6. [Medline].

  31. Millward SF, Aquino J Jr, Peterson RA. Oversized inferior vena cava: use of a single Vena Tech--LGM filter. Can Assoc Radiol J. 1996 Aug. 47(4):272-4. [Medline].

  32. Millward SF, Marsh JI, Peterson RA, et al. LGM (Vena Tech) vena cava filter: clinical experience in 64 patients. J Vasc Interv Radiol. 1991 Nov. 2(4):429-33. [Medline].

  33. Reed RA, Teitelbaum GP, Taylor FC, et al. Incomplete opening of LGM (Vena Tech) filters inserted via the transjugular approach. J Vasc Interv Radiol. 1991 Nov. 2(4):441-5. [Medline].

  34. Murphy TP, Dorfman GS, Yedlicka JW, et al. LGM vena cava filter: objective evaluation of early results. J Vasc Interv Radiol. 1991 Feb. 2(1):107-15. [Medline].

  35. Crochet DP, Brunel P, Trogrlic S, Grossetete R, Auget JL, Dary C. Long-term follow-up of Vena Tech-LGM filter: predictors and frequency of caval occlusion. J Vasc Interv Radiol. 1999 Feb. 10(2 Pt 1):137-42. [Medline].

  36. Crochet DP, Stora O, Ferry D, et al. Vena Tech-LGM filter: long-term results of a prospective study. Radiology. 1993 Sep. 188(3):857-60. [Medline].

  37. Le Blanche AF, Benazzouz A, Reynaud P, et al. The VenaTech LP permanent caval filter: effectiveness and safety in the prevention of pulmonary embolism--a European multicenter study. J Vasc Interv Radiol. 2008 Apr. 19(4):509-15. [Medline].

  38. Simon M, Kaplow R, Salzman E, Freiman D. A vena cava filter using thermal shape memory alloy. Experimental aspects. Radiology. 1977 Oct. 125(1):87-94. [Medline].

  39. Engmann E, Asch MR. Clinical experience with the antecubital Simon nitinol IVC filter. J Vasc Interv Radiol. 1998 Sep-Oct. 9(5):774-8. [Medline].

  40. Poletti PA, Becker CD, Prina L, et al. Long-term results of the Simon nitinol inferior vena cava filter. Eur Radiol. 1998. 8(2):289-94. [Medline].

  41. Aswad MA, Sandager GP, Pais SO, et al. Early duplex scan evaluation of four vena caval interruption devices. J Vasc Surg. 1996 Nov. 24(5):809-18. [Medline].

  42. Simon M, Athanasoulis CA, Kim D, et al. Simon nitinol inferior vena cava filter: initial clinical experience. Work in progress. Radiology. 1989 Jul. 172(1):99-103. [Medline].

  43. Grassi CJ, Matsumoto AH, Teitelbaum GP. Vena caval occlusion after Simon nitinol filter placement: identification with MR imaging in patients with malignancy. J Vasc Interv Radiol. 1992 Aug. 3(3):535-9. [Medline].

  44. Yallampalli S, Irani Z, Kalva SP. Endovascular removal of a permanent "TrapEase" inferior vena cava filter. Vasc Endovascular Surg. 2013 Jul. 47(5):379-82. [Medline].

  45. Davison BD, Grassi CJ. TrapEase inferior vena cava filter placed via the basilic arm vein: a new antecubital access. J Vasc Interv Radiol. 2002 Jan. 13(1):107-9. [Medline].

  46. Stone PA, Aburahma AF, Hass SM, et al. TrapEase inferior vena cava filter placement: use of the subclavian vein. Vasc Endovascular Surg. 2004 Nov-Dec. 38(6):505-9. [Medline].

  47. Rousseau H, Perreault P, Otal P, et al. The 6-F nitinol TrapEase inferior vena cava filter: results of a prospective multicenter trial. J Vasc Interv Radiol. 2001 Mar. 12(3):299-304. [Medline].

  48. Kalva SP, Wicky S, Waltman AC, Athanasoulis CA. TrapEase vena cava filter: experience in 751 patients. J Endovasc Ther. 2006 Jun. 13(3):365-72. [Medline].

  49. Bruckheimer E, Judelman AG, Bruckheimer SD, Tavori I, Naor G, Katzen BT. In vitro evaluation of a retrievable low-profile nitinol vena cava filter. J Vasc Interv Radiol. 2003 Apr. 14(4):469-74. [Medline].

  50. Kim HS, Young MJ, Narayan AK, Hong K, Liddell RP, Streiff MB. A comparison of clinical outcomes with retrievable and permanent inferior vena cava filters. J Vasc Interv Radiol. 2008 Mar. 19(3):393-9. [Medline].

  51. Berczi V, Bottomley JR, Thomas SM, Taneja S, Gaines PA, Cleveland TJ. Long-term retrievability of IVC filters: should we abandon permanent devices?. Cardiovasc Intervent Radiol. 2007 Sep-Oct. 30(5):820-7. [Medline].

  52. Gunther RW, Neuerburg J, Mossdorf A, et al. New optional IVC filter for percutaneous retrieval--in vitro evaluation of embolus capturing efficiency. Rofo. 2005 May. 177(5):632-6. [Medline].

  53. Janne d'Othee B, Faintuch S, Reedy AW, Nickerson CF, Rosen MP. Retrievable versus permanent caval filter procedures: when are they cost-effective for interventional radiology?. J Vasc Interv Radiol. 2008 Mar. 19(3):384-92. [Medline].

  54. Kaufman JA, Kinney TB, Streiff MB, et al. Guidelines for the use of retrievable and convertible vena cava filters: report from the Society of Interventional Radiology multidisciplinary consensus conference. J Vasc Interv Radiol. 2006 Mar. 17(3):449-59. [Medline].

  55. Mahrer A, Zippel D, Garniek A, et al. Retrievable vena cava filters in major trauma patients: prevalence of thrombus within the filter. Cardiovasc Intervent Radiol. 2008 Jul-Aug. 31(4):785-9. [Medline].

  56. Oliva VL, Perreault P, Giroux MF, Bouchard L, Therasse E, Soulez G. Recovery G2 inferior vena cava filter: technical success and safety of retrieval. J Vasc Interv Radiol. 2008 Jun. 19(6):884-9. [Medline].

  57. Pellerin O, Barral FG, Lions C, Novelli L, Beregi JP, Sapoval M. Early and late retrieval of the ALN removable vena cava filter: results from a multicenter study. Cardiovasc Intervent Radiol. 2008 Sep-Oct. 31(5):889-96. [Medline].

  58. Strauss EJ, Egol KA, Alaia M, Hansen D, Bashar M, Steiger D. The use of retrievable inferior vena cava filters in orthopaedic patients. J Bone Joint Surg Br. 2008 May. 90(5):662-7. [Medline].

  59. Gyang E, Zayed M, Harris EJ, Lee JT, Dalman RL, Mell MW. Factors impacting follow-up care after placement of temporary inferior vena cava filters. J Vasc Surg. 2013 Apr 12. [Medline].

  60. Decousus H, Leizorovicz A, Parent F, et al. A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis. Prévention du Risque d'Embolie Pulmonaire par Interruption Cave Study Group. N Engl J Med. 1998 Feb 12. 338(7):409-15. [Medline].

  61. Gaspard SF, Gaspard DJ. Retrievable inferior vena cava filters are rarely removed. Am Surg. 2009 May. 75(5):426-8. [Medline].

  62. Buecker A, Behrendt FF, Knuchel R, et al. Long-term retrieval of modified Gunther Tulip vena cava Filters: an animal study. Invest Radiol. 2007 Oct. 42(10):692-6. [Medline].

  63. De Gregorio MA, Gamboa P, Bonilla DL, et al. Retrieval of Gunther Tulip optional vena cava filters 30 days after implantation: a prospective clinical study. J Vasc Interv Radiol. 2006 Nov. 17(11 Pt 1):1781-9. [Medline].

  64. de Gregorio MA, Gamboa P, Gimeno MJ, et al. The Gunther Tulip retrievable filter: prolonged temporary filtration by repositioning within the inferior vena cava. J Vasc Interv Radiol. 2003 Oct. 14(10):1259-65. [Medline].

  65. Hoppe H, Nutting CW, Smouse HR, et al. Günther Tulip filter retrievability multicenter study including CT follow-up: final report. J Vasc Interv Radiol. 2006 Jun. 17(6):1017-23. [Medline].

  66. Looby S, Given MF, Geoghegan T, McErlean A, Lee MJ. Gunther Tulip retrievable inferior vena caval filters: indications, efficacy, retrieval, and complications. Cardiovasc Intervent Radiol. 2007 Jan-Feb. 30(1):59-65. [Medline].

  67. Ota S, Yamada N, Tsuji A, et al. The Gunther-Tulip retrievable IVC filter: clinical experience in 118 consecutive patients. Circ J. 2008 Feb. 72(2):287-92. [Medline].

  68. Ray CE Jr, Mitchell E, Zipser S, Kao EY, Brown CF, Moneta GL. Outcomes with retrievable inferior vena cava filters: a multicenter study. J Vasc Interv Radiol. 2006 Oct. 17(10):1595-604. [Medline].

  69. Rosenthal D, Wellons ED, Hancock SM, Burkett AB. Retrievability of the Gunther Tulip vena cava filter after dwell times longer than 180 days in patients with multiple trauma. J Endovasc Ther. 2007 Jun. 14(3):406-10. [Medline].

  70. Van Ha TG, Chien AS, Funaki BS, et al. Use of retrievable compared to permanent inferior vena cava filters: a single-institution experience. Cardiovasc Intervent Radiol. 2008 Mar-Apr. 31(2):308-15. [Medline].

  71. Wicky S, Doenz F, Meuwly JY, Portier F, Schnyder P, Denys A. Clinical experience with retrievable Günther Tulip vena cava filters. J Endovasc Ther. 2003 Oct. 10(5):994-1000. [Medline].

  72. Yamagami T, Kato T, Hirota T, Yoshimatsu R, Matsumoto T, Nishimura T. Evaluation of retrievability of the Gunther tulip vena cava filter. Cardiovasc Intervent Radiol. 2007 Mar-Apr. 30(2):226-31. [Medline].

  73. Sag AA, Stavas JM, Burke CT, Dixon RG, Marquess JS, Mauro MA. Analysis of tilt of the Gunther Tulip filter. J Vasc Interv Radiol. 2008 May. 19(5):669-76. [Medline].

  74. Smouse HB, Rosenthal D, Thuong VH, et al. Long-term retrieval success rate profile for the Günther Tulip vena cava filter. J Vasc Interv Radiol. 2009 Jul. 20(7):871-7; quiz 878. [Medline].

  75. Asch MR. Initial experience in humans with a new retrievable inferior vena cava filter. Radiology. 2002 Dec. 225(3):835-44. [Medline].

  76. Grande WJ, Trerotola SO, Reilly PM, et al. Experience with the recovery filter as a retrievable inferior vena cava filter. J Vasc Interv Radiol. 2005 Sep. 16(9):1189-93. [Medline].

  77. Kalva SP, Athanasoulis CA, Fan CM, et al. "Recovery" vena cava filter: experience in 96 patients. Cardiovasc Intervent Radiol. 2006 Jul-Aug. 29(4):559-64. [Medline].

  78. Hull JE, Robertson SW. Bard Recovery filter: evaluation and management of vena cava limb perforation, fracture, and migration. J Vasc Interv Radiol. 2009 Jan. 20(1):52-60. [Medline].

  79. Cantwell CP, Pennypacker J, Singh H, Scorza LB, Waybill PN, Lynch FC. Comparison of the recovery and G2 filter as retrievable inferior vena cava filters. J Vasc Interv Radiol. 2009 Sep. 20(9):1193-9. [Medline].

  80. Cantwell CP, Lynch FC. Caudal migration of a G2 filter during carbon dioxide cavography. J Vasc Interv Radiol. 2007 Jun. 18(6):814-5. [Medline].

  81. Oliva VL, Perreault P, Giroux MF, Bouchard L, Therasse E, Soulez G. Recovery G2 inferior vena cava filter: technical success and safety of retrieval. J Vasc Interv Radiol. 2008 Jun. 19(6):884-9. [Medline].

  82. Charles HW, Black M, Kovacs S, et al. G2 inferior vena cava filter: retrievability and safety. J Vasc Interv Radiol. 2009 Aug. 20(8):1046-51. [Medline].

  83. Binkert CA, Drooz AT, Caridi JG, et al. Technical success and safety of retrieval of the G2 filter in a prospective, multicenter study. J Vasc Interv Radiol. 2009 Nov. 20(11):1449-53. [Medline].

  84. Lynch FC, Kekulawela S. Removal of the G2 filter: differences between implantation times greater and less than 180 days. J Vasc Interv Radiol. 2009 Sep. 20(9):1200-9. [Medline].

  85. Kuo WT, Loh CT, Sze DY. Emergency retrieval of a G2 filter after complete migration into the right ventricle. J Vasc Interv Radiol. 2007 Sep. 18(9):1177-82. [Medline].

  86. Bui JT, West DL, Pinto C, Gramling-Babb P, Owens CA. Right ventricular migration and endovascular removal of an inferior vena cava filter. J Vasc Interv Radiol. 2008 Jan. 19(1):141-4. [Medline].

  87. Kim D, Schlam BW, Porter DH, Simon M. Insertion of the Simon nitinol caval filter: value of the antecubital vein approach. AJR Am J Roentgenol. 1991 Sep. 157(3):521-2. [Medline].

  88. Keller IS, Meier C, Pfiffner R, Keller E, Pfammatter T. Clinical comparison of two optional vena cava filters. J Vasc Interv Radiol. 2007 Apr. 18(4):505-11. [Medline].

  89. Oliva VL, Szatmari F, Giroux MF, Flemming BK, Cohen SA, Soulez G. The Jonas study: evaluation of the retrievability of the Cordis OptEase inferior vena cava filter. J Vasc Interv Radiol. 2005 Nov. 16(11):1439-45; quiz 1445. [Medline].

  90. Rosenthal D, Swischuk JL, Cohen SA, Wellons ED. OptEase retrievable inferior vena cava filter: initial multicenter experience. Vascular. 2005 Sep-Oct. 13(5):286-9. [Medline].

  91. Corriere MA, Sauve KJ, Ayerdi J, et al. Vena cava filters and inferior vena cava thrombosis. J Vasc Surg. 2007 Apr. 45(4):789-94. [Medline].

  92. Sadaf A, Rasuli P, Olivier A, et al. Significant caval penetration by the celect inferior vena cava filter: attributable to filter design?. J Vasc Interv Radiol. 2007 Nov. 18(11):1447-50. [Medline].

  93. Doody O, Given MF, Kavnoudias H, Street M, Thomson KR, Lyon SM. Initial experience in 115 patients with the retrievable Cook Celect vena cava filter. J Med Imaging Radiat Oncol. 2009 Feb. 53(1):64-8. [Medline].

  94. Rosenthal D, Kochupura PV, Wellons ED, Burkett AB, Methodius-Rayford WC. Günther Tulip and Celect IVC filters in multiple-trauma patients. J Endovasc Ther. 2009 Aug. 16(4):494-9. [Medline].

  95. Sangwaiya MJ, Marentis TC, Walker TG, Stecker M, Wicky ST, Kalva SP. Safety and effectiveness of the celect inferior vena cava filter: preliminary results. J Vasc Interv Radiol. 2009 Sep. 20(9):1188-92. [Medline].

  96. Lyon SM, Riojas GE, Uberoi R, et al. Short- and long-term retrievability of the Celect vena cava filter: results from a multi-institutional registry. J Vasc Interv Radiol. 2009 Nov. 20(11):1441-8. [Medline].

  97. Mismetti P, Rivron-Guillot K, Quenet S, et al. A prospective long-term study of 220 patients with a retrievable vena cava filter for secondary prevention of venous thromboembolism. Chest. 2007 Jan. 131(1):223-9. [Medline].

  98. Pacouret G, Alison D, Pottier JM, Bertrand P, Charbonnier B. Free-floating thrombus and embolic risk in patients with angiographically confirmed proximal deep venous thrombosis. A prospective study. Arch Intern Med. 1997 Feb 10. 157(3):305-8. [Medline].

  99. Moore FD Jr, Osteen RT, Karp DD, Steele G Jr, Wilson RE. Anticoagulants, venous thromboembolism, and the cancer patient. Arch Surg. 1981 Apr. 116(4):405-7. [Medline].

  100. Opinions regarding the diagnosis and management of venous thromboembolic disease. ACCP Consensus Committee on Pulmonary Embolism. American College of Chest Physicians. Chest. 1998 Feb. 113(2):499-504. [Medline].

  101. Sarosiek S, Crowther M, Sloan JM. Indications, complications, and management of inferior vena cava filters: the experience in 952 patients at an academic hospital with a level I trauma center. JAMA Intern Med. 2013 Apr 8. 173(7):513-7. [Medline].

  102. Gorman PH, Qadri SF, Rao-Patel A. Prophylactic inferior vena cava (IVC) filter placement may increase the relative risk of deep venous thrombosis after acute spinal cord injury. J Trauma. 2009 Mar. 66(3):707-12. [Medline].

  103. Gray RR, Sadler DJ, Shulman L, Saliken JC, So CB. Should anticoagulant therapy be stopped or reversed before venous intervention?. Can Assoc Radiol J. 1999 Oct. 50(5):306-9. [Medline].

  104. Aidinian G, Fox CJ, White PW, Cox MW, Adams ED, Gillespie DL. Intravascular ultrasound--guided inferior vena cava filter placement in the military multitrauma patients: a single-center experience. Vasc Endovascular Surg. 2009 Oct-Nov. 43(5):497-501. [Medline].

  105. Joels CS, Sing RF, Heniford BT. Complications of inferior vena cava filters. Am Surg. 2003 Aug. 69(8):654-9. [Medline].

  106. Arrive L, Menu Y, Dessarts I, et al. Diagnosis of abdominal venous thrombosis by means of spin-echo and gradient-echo MR imaging: analysis with receiver operating characteristic curves. Radiology. 1991 Dec. 181(3):661-8. [Medline].

  107. Ascer E, Gennaro M, Lorensen E, Pollina RM. Superior vena caval Greenfield filters: indications, techniques, and results. J Vasc Surg. 1996 Mar. 23(3):498-503. [Medline].

  108. Awh MH, Taylor FC, Lu CT. Spontaneous fracture of a Vena-Tech inferior vena caval filter. AJR Am J Roentgenol. 1991 Jul. 157(1):177-8. [Medline].

  109. Barth KH, Matsumoto AH. Patient care in interventional radiology: a perspective. Radiology. 1991 Jan. 178(1):11-7. [Medline].

  110. Benjamin ME, Sandager GP, Cohn EJ Jr, et al. Duplex ultrasound insertion of inferior vena cava filters in multitrauma patients. Am J Surg. 1999 Aug. 178(2):92-7. [Medline].

  111. Bonn J, Liu JB, Eschelman DJ, Sullivan KL, Pinheiro LW, Gardiner GA Jr. Intravascular ultrasound as an alternative to positive-contrast vena cavography prior to filter placement. J Vasc Interv Radiol. 1999 Jul-Aug. 10(7):843-9. [Medline].

  112. Bonnichon P, Gaudard F, Lecam B, et al. Biometry of the infrarenal inferior vena cava measured by computed tomography. Clinical applications. Surg Radiol Anat. 1992. 14(3):265-9. [Medline].

  113. Boyd-Kranis R, Sullivan KL, Eschelman DJ, Bonn J, Gardiner GA. Accuracy and safety of carbon dioxide inferior vena cavography. J Vasc Interv Radiol. 1999 Oct. 10(9):1183-9. [Medline].

  114. Brasel KJ, Borgstrom DC, Weigelt JA. Cost-effective prevention of pulmonary embolus in high-risk trauma patients. J Trauma. 1997 Mar. 42(3):456-60; discussion 460-2. [Medline].

  115. Brown DB, Labuski MR, Cardella JF, Singh H, Waybill PN. Determination of inferior vena cava diameter in the angiography suite: comparison of three common methods. J Vasc Interv Radiol. 1999 Feb. 10(2 Pt 1):143-7. [Medline].

  116. Browne RJ, Estrada FP. Guidewire entrapment during Greenfield filter deployment. J Vasc Surg. 1998 Jan. 27(1):174-6. [Medline].

  117. Cho KJ, Greenfield LJ, Proctor MC, et al. Evaluation of a new percutaneous stainless steel Greenfield filter. J Vasc Interv Radiol. 1997 Mar-Apr. 8(2):181-7. [Medline].

  118. Cohen JR, Grella L, Citron M. Greenfield filter instead of heparin as primary treatment for deep venous thrombosis or pulmonary embolism in patients with cancer. Cancer. 1992 Oct 1. 70(7):1993-6. [Medline].

  119. Couch GG, Johnston KW, Ojha M. An in vitro comparison of the hemodynamics of two inferior vena cava filters. J Vasc Surg. 2000 Mar. 31(3):539-49. [Medline].

  120. Cronan JJ. Venous thromboembolic disease: the role of US. Radiology. 1993 Mar. 186(3):619-30. [Medline].

  121. Dennis JW, Menawat S, Von Thron J, et al. Efficacy of deep venous thrombosis prophylaxis in trauma patients and identification of high-risk groups. J Trauma. 1993 Jul. 35(1):132-8; discussion 138-9. [Medline].

  122. Domingo ML, Marti-Bonmati L, Dosda R, Pallardo Y. Interobserver agreement in the diagnosis of pulmonary embolism with helical CT. Eur J Radiol. 2000 May. 34(2):136-40. [Medline].

  123. Dorfman GS. Percutaneous inferior vena caval filters. Radiology. 1990 Mar. 174(3 Pt 2):987-92. [Medline].

  124. Drucker EA, Rivitz SM, Shepard JA, et al. Acute pulmonary embolism: assessment of helical CT for diagnosis. Radiology. 1998 Oct. 209(1):235-41. [Medline].

  125. Ferris DJ, Shah HR. The inferior vena cava. Baum S, Pentecost MJ, eds. Abrams' Angiography. 4th ed. Boston: Little Brown & Co; 1997. Vol 2: 932-74.

  126. Ferris EJ, McCowan TC, Carver DK, McFarland DR. Percutaneous inferior vena caval filters: follow-up of seven designs in 320 patients. Radiology. 1993 Sep. 188(3):851-6. [Medline].

  127. Friedland M, Kazmers A, Kline R, et al. Vena cava duplex imaging before caval interruption. J Vasc Surg. 1996 Oct. 24(4):608-12; discussion 612-3. [Medline].

  128. Goldhaber SZ. Pulmonary embolism. N Engl J Med. 1998 Jul 9. 339(2):93-104. [Medline].

  129. Goodman LR, Lipchik RJ. Diagnosis of acute pulmonary embolism: time for a new approach. Radiology. 1996 Apr. 199(1):25-7. [Medline].

  130. Goodman LR, Lipchik RJ, Kuzo RS, Liu Y, McAuliffe TL, O'Brien DJ. Subsequent pulmonary embolism: risk after a negative helical CT pulmonary angiogram--prospective comparison with scintigraphy. Radiology. 2000 May. 215(2):535-42. [Medline].

  131. Gordon AC, Saliken JC, Johns D, Owen R, Gray RR. US-guided puncture of the internal jugular vein: complications and anatomic considerations. J Vasc Interv Radiol. 1998 Mar-Apr. 9(2):333-8. [Medline].

  132. Gosin JS, Graham AM, Ciocca RG, Hammond JS. Efficacy of prophylactic vena cava filters in high-risk trauma patients. Ann Vasc Surg. 1997 Jan. 11(1):100-5. [Medline].

  133. Grassi CJ, Swan TL, Cardella JF, et al. Quality improvement guidelines for percutaneous permanent inferior vena cava filter placement for the prevention of pulmonary embolism. J Vasc Interv Radiol. 2003 Sep. 14(9 Pt 2):S271-5. [Medline].

  134. Greenfield LJ. The PREPIC Study Group. Eight-year follow-up of patients with permanent vena cava filters in the prevention of pulmonary embolism: the PREPIC (Prevention du Risque d'Embolie Pulmonaire par Interruption Cave) Randomized Study. Perspect Vasc Surg Endovasc Ther. 2006 Jun. 18(2):187-8. [Medline].

  135. Greenfield LJ, Cho KJ, Proctor MC, Sobel M, Shah S, Wingo J. Late results of suprarenal Greenfield vena cava filter placement. Arch Surg. 1992 Aug. 127(8):969-73. [Medline].

  136. Greenfield LJ, Proctor MC. Experimental embolic capture by asymmetric Greenfield filters. J Vasc Surg. 1992 Sep. 16(3):436-43; discussion 443-4. [Medline].

  137. Greenfield LJ, Proctor MC. Recurrent thromboembolism in patients with vena cava filters. J Vasc Surg. 2001 Mar. 33(3):510-4. [Medline].

  138. Greenfield LJ, Proctor MC. Suprarenal filter placement. J Vasc Surg. 1998 Sep. 28(3):432-8; discussion 438. [Medline].

  139. Greenfield LJ, Proctor MC, Saluja A. Clinical results of Greenfield filter use in patients with cancer. Cardiovasc Surg. 1997 Apr. 5(2):145-9. [Medline].

  140. Hammer FD, Rousseau HP, Joffre FG, Sentenac BP, Tran-Van T, Barthelemy RP. In vitro evaluation of vena cava filters. J Vasc Interv Radiol. 1994 Nov-Dec. 5(6):869-76. [Medline].

  141. Harker LA. Antithrombotic therapy. Goldman L, Bennett JC, eds. Cecil Textbook of Medicine. 21st ed. WB Saunders Co; 2000. 1021-8.

  142. Hirsch DR, Ingenito EP, Goldhaber SZ. Prevalence of deep venous thrombosis among patients in medical intensive care. JAMA. 1995 Jul 26. 274(4):335-7. [Medline].

  143. Hoffmann U, Schima W, Herold C. Pulmonary magnetic resonance angiography. Eur Radiol. 1999. 9(9):1745-54. [Medline].

  144. Hull RD. Peripheral vascular disease. Goldman L, Bennett JC, eds. Cecil Textbook of Medicine. 21st ed. WB Saunders Co; 2000. 367-72.

  145. Kaufman JA, Thomas JW, Geller SC, Rivitz SM, Waltman AC. Guide-wire entrapment by inferior vena caval filters: in vitro evaluation. Radiology. 1996 Jan. 198(1):71-6. [Medline].

  146. Kaufman JA, Waltman AC, Rivitz SM, Geller SG. Anatomical observations on the renal veins and inferior vena cava at magnetic resonance angiography. Cardiovasc Intervent Radiol. 1995 May-Jun. 18(3):153-7. [Medline].

  147. Kaura DR, Gray RR, Sadler DJ, So CB, Saliken JC. Value of frontal caval measurement in the placement of inferior vena cava filters. Can Assoc Radiol J. 1999 Oct. 50(5):301-5. [Medline].

  148. Khansarinia S, Dennis JW, Veldenz HC, Butcher JL, Hartland L. Prophylactic Greenfield filter placement in selected high-risk trauma patients. J Vasc Surg. 1995 Sep. 22(3):231-5; discussion 235-6. [Medline].

  149. Kiproff PM, Deeb ZL, Contractor FM, Khoury MB. Magnetic resonance characteristics of the LGM vena cava filter: technical note. Cardiovasc Intervent Radiol. 1991 Jul-Aug. 14(4):254-5. [Medline].

  150. Klatsky AL, Armstrong MA, Poggi J. Risk of pulmonary embolism and/or deep venous thrombosis in Asian-Americans. Am J Cardiol. 2000 Jun 1. 85(11):1334-7. [Medline].

  151. Ko SH, Reynolds BR, Nicholas DH, et al. Institutional protocol improves retrievable inferior vena cava filter recovery rate. Surgery. 2009 Oct. 146(4):809-14; discussion 814-6. [Medline].

  152. Korbin CD, Reed RA, Taylor FC, Kokoris SP, Teitelbaum GP. In vitro flow phantom analysis and clot-capturing ability of incompletely opened Vena Tech-LGM vena caval filters. Cardiovasc Intervent Radiol. 1993 Jan-Feb. 16(1):3-6. [Medline].

  153. Langan EM 3rd, Miller RS, Casey WJ 3rd, Carsten CG 3rd, Graham RM, Taylor SM. Prophylactic inferior vena cava filters in trauma patients at high risk: follow-up examination and risk/benefit assessment. J Vasc Surg. 1999 Sep. 30(3):484-88. [Medline].

  154. LaPlante JS, Contractor FM, Kiproff PM, Khoury MB. Migration of the Simon nitinol vena cava filter to the chest. AJR Am J Roentgenol. 1993 Feb. 160(2):385-6. [Medline].

  155. Leask RL, Johnston KW, Ojha M. Hemodynamic effects of clot entrapment in the TrapEase inferior vena cava filter. J Vasc Interv Radiol. 2004 May. 15(5):485-90. [Medline].

  156. Levine MN, Gent M, Hirsh J, et al. Ardeparin (low-molecular-weight heparin) vs graduated compression stockings for the prevention of venous thromboembolism. A randomized trial in patients undergoing knee surgery. Arch Intern Med. 1996 Apr 22. 156(8):851-6. [Medline].

  157. Lewis BD, James EM, Welch TJ, Joyce JW, Hallett JW, Weaver AL. Diagnosis of acute deep venous thrombosis of the lower extremities: prospective evaluation of color Doppler flow imaging versus venography. Radiology. 1994 Sep. 192(3):651-5. [Medline].

  158. Lynch FC. Balloon-assisted removal of tilted inferior vena cava filters with embedded tips. J Vasc Interv Radiol. 2009 Sep. 20(9):1210-4. [Medline].

  159. Matchett WJ, Jones MP, McFarland DR, Ferris EJ. Suprarenal vena caval filter placement: follow-up of four filter types in 22 patients. J Vasc Interv Radiol. 1998 Jul-Aug. 9(4):588-93. [Medline].

  160. Mayo JR, Remy-Jardin M, Muller NL, et al. Pulmonary embolism: prospective comparison of spiral CT with ventilation-perfusion scintigraphy. Radiology. 1997 Nov. 205(2):447-52. [Medline].

  161. McCowan TC, Ferris EJ, Carver DK. Inferior vena caval filter thrombi: evaluation with intravascular US. Radiology. 1990 Dec. 177(3):783-8. [Medline].

  162. McCowan TC, Ferris EJ, Carver DK, Harshfield DL. Use of external jugular vein as a route for percutaneous inferior vena caval filter placement. Radiology. 1990 Aug. 176(2):527-30. [Medline].

  163. McCowan TC, Ferris EJ, Carver DK, Molpus WM. Complications of the nitinol vena caval filter. J Vasc Interv Radiol. 1992 May. 3(2):401-8. [Medline].

  164. McMurtry AL, Owings JT, Anderson JT, Battistella FD, Gosselin R. Increased use of prophylactic vena cava filters in trauma patients failed to decrease overall incidence of pulmonary embolism. J Am Coll Surg. 1999 Sep. 189(3):314-20. [Medline].

  165. Menzoian JO, LoGerfo FW, Doyle JE, Weitzman AF, Sequeira JC. Technical modifications in the placement of inferior vena caval filter devices. Am J Surg. 1981 Aug. 142(2):216-8. [Medline].

  166. Nunn CR, Neuzil D, Naslund T, et al. Cost-effective method for bedside insertion of vena caval filters in trauma patients. J Trauma. 1997 Nov. 43(5):752-8. [Medline].

  167. O'Malley KF, Ross SE. Pulmonary embolism in major trauma patients. J Trauma. 1990 Jun. 30(6):748-50. [Medline].

  168. Oppat WF, Chiou AC, Matsumura JS. Intravascular ultrasound-guided vena cava filter placement. J Endovasc Surg. 1999 Aug. 6(3):285-7. [Medline].

  169. Pellerin O, Barral FG, Lions C, Novelli L, Beregi JP, Sapoval M. Early and late retrieval of the ALN removable vena cava filter: results from a multicenter study. Cardiovasc Intervent Radiol. 2008 Sep-Oct. 31(5):889-96. [Medline].

  170. Pena CS, Fan CM, Geller SC. Initial experiences with the Trapease inferior vena cava filter. J Vasc Interv Radiol. 2001. 12:S139.

  171. Quinn DA, Thompson BT, Terrin ML, et al. A prospective investigation of pulmonary embolism in women and men. JAMA. 1992 Oct 7. 268(13):1689-96. [Medline].

  172. Remy-Jardin M, Remy J. Spiral CT angiography of the pulmonary circulation. Radiology. 1999 Sep. 212(3):615-36. [Medline].

  173. Remy-Jardin M, Remy J, Artaud D, Fribourg M, Beregi JP. Spiral CT of pulmonary embolism: diagnostic approach, interpretive pitfalls and current indications. Eur Radiol. 1998. 8(8):1376-90. [Medline].

  174. Remy-Jardin M, Remy J, Wattinne L, Giraud F. Central pulmonary thromboembolism: diagnosis with spiral volumetric CT with the single-breath-hold technique--comparison with pulmonary angiography. Radiology. 1992 Nov. 185(2):381-7. [Medline].

  175. Ricco JB, Dubreuil F, Reynaud P, et al. The LGM Vena-Tech caval filter: results of a multicenter study. Ann Vasc Surg. 1995. 9 Suppl:S89-100. [Medline].

  176. Roditi GH, Buff BL, Longmaid HE. MR venography of left renal vein anomalies. Clin Radiol. 1996 Dec. 51(12):861-4. [Medline].

  177. Rodriguez JL, Lopez JM, Proctor MC, et al. Early placement of prophylactic vena caval filters in injured patients at high risk for pulmonary embolism. J Trauma. 1996 May. 40(5):797-802; discussion 802-4. [Medline].

  178. Rogers FB, Shackford SR, Ricci MA, Huber BM, Atkins T. Prophylactic vena cava filter insertion in selected high-risk orthopaedic trauma patients. J Orthop Trauma. 1997 May. 11(4):267-72. [Medline].

  179. Rogers FB, Shackford SR, Wilson J, Ricci MA, Morris CS. Prophylactic vena cava filter insertion in severely injured trauma patients: indications and preliminary results. J Trauma. 1993 Oct. 35(4):637-41; discussion 641-2. [Medline].

  180. Rogoff PA, Hilgenberg AD, Miller SL, Stephan SM. Cephalic migration of the bird's nest inferior vena caval filter: report of two cases. Radiology. 1992 Sep. 184(3):819-22. [Medline].

  181. Rosenthal D, Wellons ED, Levitt AB, Shuler FW, O'Conner RE, Henderson VJ. Role of prophylactic temporary inferior vena cava filters placed at the ICU bedside under intravascular ultrasound guidance in patients with multiple trauma. J Vasc Surg. 2004 Nov. 40(5):958-64. [Medline].

  182. Rousseau HP, Perreault P, Otal P, et al. First clinical experience with a new 6F nitinol inferior vena cava filter: results from a European multicenter trial. J Vasc Interv Radiol. 2000. 11:196.

  183. Ruiz AJ, Hill SL, Berry RE. Heparin, deep venous thrombosis, and trauma patients. Am J Surg. 1991 Aug. 162(2):159-62. [Medline].

  184. Sato DT, Robinson KD, Gregory RT, et al. Duplex directed caval filter insertion in multi-trauma and critically ill patients. Ann Vasc Surg. 1999 Jul. 13(4):365-71. [Medline].

  185. Savader SJ, Gomez-Jorge J. Peripheral and central deep vein thrombosis. Savader SJ, Trerotola SO, eds. Venous Interventional Radiology With Clinical Perspectives. 2nd ed. New York, NY: Thieme; 2000. 431-57.

  186. Savader SJ, Ronsivalle JA. Permanent inferior vena cava filter. Savader SJ, Trerotola SO, eds. Venous Interventional Radiology With Clinical Perspectives. 2nd ed. New York, NY: Thieme; 2000. 511-27.

  187. Schanzer H, Schanzer A. Guidewire entrapment during deployment of the over-the-guidewire stainless steel Greenfield filter: a device design-related complication. J Vasc Surg. 2000 Mar. 31(3):607-10. [Medline].

  188. Schutzer R, Ascher E, Hingorani A, Jacob T, Kallakuri S. Preliminary results of the new 6F TrapEase inferior vena cava filter. Ann Vasc Surg. 2003 Jan. 17(1):103-6. [Medline].

  189. Shi HJ, Huang YH, Shen T, Xu Q. Percutaneous mechanical thrombectomy combined with catheter-directed thrombolysis in the treatment of symptomatic lower extremity deep venous thrombosis. Eur J Radiol. 2009 Aug. 71(2):350-5. [Medline].

  190. Silverstein MD, Heit JA, Mohr DN, Petterson TM, O'Fallon WM, Melton LJ 3rd. Trends in the incidence of deep vein thrombosis and pulmonary embolism: a 25-year population-based study. Arch Intern Med. 1998 Mar 23. 158(6):585-93. [Medline].

  191. Simon M, Rabkin DJ, Kleshinski S, Kim D, Ransil BJ. Comparative evaluation of clinically available inferior vena cava filters with an in vitro physiologic simulation of the vena cava. Radiology. 1993 Dec. 189(3):769-74. [Medline].

  192. Sing RF, Cicci CK, Smith CH, Messick WJ. Bedside insertion of inferior vena cava filters in the intensive care unit. J Trauma. 1999 Dec. 47(6):1104-7. [Medline].

  193. Smouse HB, Rosenthal D, Thuong VH, et al. Long-term retrieval success rate profile for the Günther Tulip vena cava filter. J Vasc Interv Radiol. 2009 Jul. 20(7):871-7; quiz 878. [Medline].

  194. Smouse HB, Van Alstine WG, Mack S, McCann-Brown JA. Deployment performance and retrievability of the Cook Celect vena cava filter. J Vasc Interv Radiol. 2009 Mar. 20(3):375-83. [Medline].

  195. Soler R, Rodriguez E, Lopez MF, Marini M. MR imaging in inferior vena cava thrombosis. Eur J Radiol. 1995 Jan. 19(2):101-7. [Medline].

  196. Spain DA, Richardson JD, Polk HC Jr, Bergamini TM, Wilson MA, Miller FB. Venous thromboembolism in the high-risk trauma patient: do risks justify aggressive screening and prophylaxis?. J Trauma. 1997 Mar. 42(3):463-7; discussion 467-9. [Medline].

  197. Spence LD, Gironta MG, Malde HM, Mickolick CT, Geisinger MA, Dolmatch BL. Acute upper extremity deep venous thrombosis: safety and effectiveness of superior vena caval filters. Radiology. 1999 Jan. 210(1):53-8. [Medline].

  198. Spencer FC, Quattlebaum JK, Quattlebaum JK Jr, Sharp EH, Jude JR. Plication of the inferior vena cava for pulmonary embolism: a report of 20 cases. Ann Surg. 1962 Jun. 155:827-37. [Medline]. [Full Text].

  199. Stein PD, Alnas M, Skaf E, et al. Outcome and complications of retrievable inferior vena cava filters. Am J Cardiol. 2004 Oct 15. 94(8):1090-3. [Medline].

  200. Taylor FC, Awh MH, Kahn CE Jr, Lu CT. Vena Tech vena cava filter: experience and early follow-up. J Vasc Interv Radiol. 1991 Nov. 2(4):435-40. [Medline].

  201. Terhaar OA, Lyon SM, Given MF, Foster AE, Mc Grath F, Lee MJ. Extended interval for retrieval of Günther Tulip filters. J Vasc Interv Radiol. 2004 Nov. 15(11):1257-62. [Medline].

  202. Tola JC, Holtzman R, Lottenberg L. Bedside placement of inferior vena cava filters in the intensive care unit. Am Surg. 1999 Sep. 65(9):833-7; discussion 837-8. [Medline].

  203. van Beek EJ, Kuijer PM, Buller HR, Brandjes DP, Bossuyt PM, ten Cate JW. The clinical course of patients with suspected pulmonary embolism. Arch Intern Med. 1997 Dec 8-22. 157(22):2593-8. [Medline].

  204. Van Natta TL, Morris JA Jr, Eddy VA, et al. Elective bedside surgery in critically injured patients is safe and cost-effective. Ann Surg. 1998 May. 227(5):618-24; discussion 624-6. [Medline]. [Full Text].

  205. van Rossum AB, Pattynama PM, Mallens WM, Hermans J, Heijerman HG. Can helical CT replace scintigraphy in the diagnostic process in suspected pulmonary embolism? A retrolective-prolective cohort study focusing on total diagnostic yield. Eur Radiol. 1998. 8(1):90-6. [Medline].

  206. Vesely T, Darcy M, Picus D, Hicks M. Technical problems associated with placement of the Bird's Nest inferior vena cava filter. AJR Am J Roentgenol. 1992 Apr. 158(4):875-80. [Medline].

  207. Vrachliotis TG, Rabkin DJ, Lang EV. In vitro capturing characteristics of the TrapEase filter. J Vasc Interv Radiol. 2001. 12:S78.

  208. Warkentin TE, Sikov WM, Lillicrap DP. Multicentric warfarin-induced skin necrosis complicating heparin-induced thrombocytopenia. Am J Hematol. 1999 Sep. 62(1):44-8. [Medline].

  209. West DL, Bui JT, Hanhan S, Owens CA. Percutaneous retrieval of a vena tech LP filter after failure of filter opening. J Vasc Interv Radiol. 2006 Apr. 17(4):739-41. [Medline].

  210. White KE, McLean GK. Bird's nest filter: inferior strut migration during massive thromboembolization. J Vasc Interv Radiol. 1996 Jul-Aug. 7(4):537-40. [Medline].

  211. Wittenberg G, Kueppers V, Tschammler A, Scheppach W, Kenn W, Hahn D. Long-term results of vena cava filters: experiences with the LGM and the Titanium Greenfield devices. Cardiovasc Intervent Radiol. 1998 May-Jun. 21(3):225-9. [Medline].

  212. Woodard PK. CT scan negative for pulmonary embolism: where do we go from here?. Radiology. 2000 May. 215(2):325-6. [Medline].

  213. Yavuz K, Geyik S, Hoppe H, Kolbeck KJ, Kaufman JA. Venous thromboembolism after retrieval of inferior vena cava filters. J Vasc Interv Radiol. 2008 Apr. 19(4):504-508. [Medline].

  214. Zeni PT Jr, Roberts JA, Price M, McFarland DR, Ferris EJ. Titanium Greenfield filter placement: a retrospective analysis of filter tilting comparing access by right internal jugular vein with right femoral vein. AJR Am J Roentgenol. 1997 Sep. 169(3):739-41. [Medline].

  215. Bartlett MA, Mauck KF, Daniels PR. Prevention of venous thromboembolism in patients undergoing bariatric surgery. Vasc Health Risk Manag. 2015. 11:461-77. [Medline].

 
Previous
Next
 
Anteroposterior image from an inferior venacavographic examination demonstrates an inferior vena cava of normal diameter without thrombus. In addition, the inflow from the renal veins can be seen at the level of the L1 vertebral body.
Anteroposterior image from an inferior venacavographic examination demonstrates duplication of the inferior vena cava. The left-sided inferior vena cava drains into the left renal vein.
Anteroposterior image from an inferior venacavographic examination demonstrates a circumaortic left renal vein.
Bird's nest filter.
Greenfield filter.
Simon nitinol filter.
TrapEase filter.
 
 
 