Diagnostic Considerations
Pseudomalfunction
In pseudomalfunction of the pacemaker, although an apparent pacing system malfunction is suggested clinically, the apparent malfunction is a normal programmed pacer function. This is partly due to new algorithms (eg, Managed Ventricular Pacing [MVP]) [9] to preserve intrinsic conduction and more physiologic pacing.
Examples of pseudomalfunction include the following:
-
Functional undersense with magnet application, safety pacing, triggered mode, fusion and pseudofusion beats, and blanking period
-
Functional oversense with short ventricular blanking period
-
Inappropriate rate with rate hysteresis, rate-drop response, rate smoothing, rate-responsive mode, programmed rest/sleep rate, and mode switch
-
Inappropriate mode with magnet application and mode switch
Manufacturers' advisories
Manufacturer recalls or advisories due to unanticipated device malfunction after their release may require replacement of the pulse generator, lead, or both. A decision analysis model by Amin et al warrants device replacement in pacemaker-dependent patients if a manufacturer advisory device failure rate is over 0.3%. [10]
Differential Diagnoses
-
Acute heart failure
-
Pacemaker Malfunction. Atrial undersensing. The rhythm strip shows an atrial pacing artifact after the intrinsic P wave.
-
Pacemaker Malfunction. Ventricular undersensing. The rhythm strip shows ventricular pacing artifacts despite normal underlying ventricular activity.
-
Pacemaker Malfunction. Atrial lead dislodgment. The chest radiograph film detail shows a dislodged atrial lead with the tip in the right ventricular cavity.
-
Pacemaker Malfunction. Ventricular noncapture. The rhythm strip shows atrial (P wave) sensing followed by a ventricular spike, which failed to capture the ventricle.
-
Pacemaker Malfunction. Loss of atrial capture. The rhythm strip shows intermittent loss of atrial capture.
-
Pacemaker Malfunction. Pacemaker-mediated tachycardia. The rhythm strip shows ventricular pacing at 110 beats per minute (programmed maximal track rate).
-
Pacemaker Malfunction. Termination of pacemaker-mediated tachycardia (PMT). Automatic postventricular atrial refractory period (PVARP) extension terminated the PMT.
-
Pacemaker Malfunction. This image shows an artifact due to monitor malfunction or a loose limb lead connection. An abrupt loss of a portion of the QRS complex followed by a flat line can be observed. If R-R intervals are matched, two QRS complexes are missing during the pause. If the artifact is due to a dislodged lead, a pacing artifact with no capture should be observed.
-
Pacemaker Malfunction. This is a typical example of ventricular oversensing with inhibition of ventricular pacing. In ventricular noncapture, a ventricular pacing artifact should be present after the third P wave.