Diaphragm Pacing Technique

Updated: Jan 22, 2018
  • Author: Shabir Bhimji, MD, PhD; Chief Editor: Zab Mosenifar, MD, FACP, FCCP  more...
  • Print
Technique

Placement

The surgical procedure for inserting the diaphragmatic pacemaker is not complicated. To stimulate the phrenic nerve, one may place the electrode in the neck or in the chest. With the neck approach, a small supraclavicular incision is made just above the clavicle. The sternocleidomastoid muscle is retracted and the phrenic nerve is identified as it courses between the middle and anterior scalene muscles. The identification of the nerve can be confirmed with a nerve simulator and use of intraoperative fluoroscopy. Stimulation of the phrenic nerve in the chest is done via a thoracoscopic approach. In the right chest cavity, the phrenic nerve can be found coursing just posterior to the esophagus; in the left chest, it is located lateral to the pericardium. The mediastinal pleura is incised parallel to the phrenic nerve, but a few millimeters of tissue are left on either side of the nerve. This incision is made both anterior and posterior to the nerve.

The diaphragm may also be stimulated directly with electrodes if necessary. Before implanting electrodes in the diaphragm, motor mapping is done and a grid map algorithm is created. Once the points of maximal contraction are identified, the electrodes are implanted into these sites and again hooked to an external stimulator. Laparoscopy is often used to implant  electrodes bilateraly on the diaphragm. [1, 2, 8, 9]

Excellent contractions of the diaphragm with pacing should be confirmed with imaging, and the threshold should be in the range of 1-2 mA. Failure to pace or the presence of a high threshold should lead to reassessment to detect reasons for lack of optimal functioning, such as improper connections, wire injury, lead dislodgment, or interposition of excess tissues between the electrode and the nerve.

The current is set on the basis of fluoroscopic testing. In most cases, a standardized method is developed to assess the optimum diaphragmatic excursion, and this system is followed each time. Maximal voluntary descent of the diaphragm during inspiration is determined for each patient. The amount of current required to generate this descent is assessed, and the current is kept at the minimum setting that produces the desired result. [1, 2, 8]

Prophylactic antibiotics should be given preoperatively to prevent infection.

Next:

Conduct of Pacing

Unlike cardiac pacemakers, diaphragmatic pacemakers cannot be used immediately. Pacing is usually not started for 2-3 weeks after implantation is complete. Early pacing usually leads to development of pleural effusions, most likely from the breakage of lung and chest wall adhesions that bleed. A program of gradual conditioning is started to allow the diaphragm to regain its muscle strength and bulk. The adaptive process should be slow and gradual. The rest period is gradually and progressively shortened until full-time pacing is achieved.

Careful assessment of lung function and arterial blood gases to look for fatigue may be advisable. At the same time, the frequency of stimuli in the pulse train is decreased progressively to minimize fatigue of the diaphragm and allow longer pacing periods. Concurrently, the respiratory rate is gradually decreased to the lowest number that provides adequate ventilation. To ensure that the adequacy of ventilation is maintained in the sitting position, a frim abdominal binder is used for quadriplegics. In addition, the respiratory rate is increased gradually until the patient appears comfortable and not hypoxic. [1, 2]

Other monitoring measures that are frequently employed include pulse oximetry, pulmonary function testing, and periodic arterial blood gas measurement. In addition, there is no substitute for a thorough physical examination. These patients need hourly monitoring during the initial phase of conditioning until pacing is advanced and tolerated. Noninvasive oxygen monitoring is routinely performed, and carbon dioxide is also monitored to ensure adequate ventilation.

If, at any time during the conditioning phase, tidal volume falls or carbon dioxide tension starts to rise, diaphragmatic fatigue should be suspected. In such cases, the patient should be allowed to rest on a mechanical ventilator for a minimum of 12 hours or, if that is not feasible, at least overnight. The next day, pacing should resume at a shorter duration.

Patients with central alveolar hypoventilation usually need to be paced at night. The daytime break often gives the diaphragm enough time to recover from fatigue. [8, 9]

Close monitoring is required during the initial conditioning period. Even after discharge, monitoring continues to be necessary because the system is still prone to errors and fatigue. Patient factors that influence pacing include hormonal changes, infection, fever, and stress, as well as any other factors that place additional stress on the diaphragm and thereby lead to extra ventilatory requirements. It is important to remember that when mechanical ventilation is started in these patients, it can lead to disuse of the diaphragm. [1]

Previous
Next:

Tracheostomy

A permanent tracheostomy is highly recommended as a safety measure for all patients who undergo diaphragm pacing. Diaphragm pacing is known to be associated with intermittent upper airway obstruction resulting from strong diaphragmatic contractions that are not synchronized with the muscles of the upper airway. At night, all patients are told to leave the tracheostomy stoma open. Thus, if periodic dysfunction occurs at night, the tracheostomy offers secure access to the airways for the institution of positive-pressure ventilation. Once full-time ventilation is achieved, the stoma can be closed with a button instead of a tracheostomy tube. The stoma button not only is acceptable cosmetically but also allows normal speech and reduces tracheal irritation and injury. [1, 2, 5, 29]

Previous
Next:

Approach Considerations

Some surgeons prefer the cervical approach because it avoids the morbidity associated with a thoracic procedure. Many of these patients have marginal pulmonary function as well as other comorbidities. The disadvantage of the cervical approach is that in a small number of patients, the current amplitude required to stimulate the phrenic nerve may also stimulate adjacent nerves in the area. For example, it has been reported that some people with cervical phrenic nerve pacers can develop stimulation of the brachial plexus, which may cause involuntary movements of the upper arm.

The benefit of the thoracic approach is that stimulation of the brachial plexus is avoided. Furthermore, there is evidence that an accessory branch of the phrenic nerve reconnects to the main trunk in the chest and that this results in good diaphragmatic stimulation. For patients who are critically ill or not able to tolerate a thoracotomy, a thoracoscopic approach is recommended.

In patients who have a paralyzed diaphragm or those who have not utilized the diaphragm for 6 or more months, conditioning of the diaphragm is required. There is no universal approach for diaphragmatic conditioning that is effective in all patients, with the duration varying from patient to patient.

In general, the following steps are undertaken to condition the diaphragm:

  • Gradually increase the duration of diaphragm pacing each day or week. Allow for rest at night, when the patient is on a mechanical ventilator.
  • During pacing, monitor the patient's arterial blood gas and pulse oximetry to assess for oxygenation and ventilation.
  • Avoid hypercapnia; if it occurs, increase the ventilatory rate.
  • The manufacturer of each device provides guidelines on the conditioning process, and their recommendations should be followed.

Complications

Complications of diaphragmatic pacing may occur from the surgical procedure or may be pacing related. They include the following:

  • Skin infection after surgery

  • Lung complications after a thoracic procedure (eg, atelectasis, pneumonia, pain)

  • Dislodgement of electrode

  • Transmission of electrical impulses to the brachial plexus, with involuntary arm movements

  • Hardware malfunction

Previous