Pulse Oximetry 

Updated: Sep 02, 2015
Author: Bruce M Lo, MD, MBA, CPE, RDMS, FACEP, FAAEM, FACHE; Chief Editor: Zab Mosenifar, MD, FACP, FCCP 

Overview

Background

Pulse oximetry is a noninvasive method of measuring the oxygenation level in the blood.

Modern pulse oximeters measure the amount of red and infrared light in an area of pulsatile blood flow. Because red light is primarily absorbed by deoxygenated blood and infrared light is primarily absorbed by oxygenated blood, the ratio of absorption can be measured. Because the amount of light absorbed varies with each pulse wave, the difference of measurement between two points in the pulse wave occurs in the arterial blood flow, with more than several hundred measurements per second. This is compared against baseline values, giving both the pulse oximetry oxygen saturation (SpO2) and the pulse rate.

Indications

Indications for pulse oximetry include the following:

  • Endotracheal intubation

  • Cardiac arrest

  • Procedural sedation

  • Asthma/chronic obstructive pulmonary disease (COPD)

  • Respiratory complaints

  • Acute respiratory distress syndrome (ARDS)

  • Sleep disorders/sleep apnea

  • Shunts in cyanotic heart diseases

Technical Considerations

Pulse oximetry probes consist of either transmission probes or reflectance probes. With transmission probes, the light emitter and sensor are placed opposite each other on pulsatile tissue such as a digit or ear. With reflectance probes, the light emitter and sensor are placed side by side on a flat body surface.

Anything that interferes with the transmission or absorbance of light can cause errors in SpO2 readings. This can be seen with a poor-quality plethysmographic tracing, suggesting possible errors in SpO2 readings.

Erroneous readings

Several situations can cause an erroneous SpO2 reading, especially with the use of transmission probes. Darker skin pigments, certain nail polishes, dyshemoglobinemias (eg, carboxyhemoglobin, methemoglobin), intravenous dyes (eg, methylene blue), hypoperfusion, and hypoxia (especially with SpO2 readings< 80%) can cause errors. Motion and exposure to ambient or excessive light has also been shown to cause erroneous SpO2 readings.[1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13]

Delay in change

SpO2 readings in distal extremities may be delayed. Compared with measurements from the earlobe, finger measurements were delayed by around 30 seconds, whereas toe measurements were delayed by up to 90 seconds.[14, 15, 16] Thus, caution must be used when interpreting SpO2 during rapid changes in oxygenation levels.

Forehead probes

Reflectance probes must be used on the forehead for reliable readings. To prevent venous pulsation from causing erroneous readings, a headband with slight pressure should be placed. Venous pooling can be also caused by placing patients in the Trendelenburg position, resulting in inaccurate SpO2 readings. The probe should be placed over a pulsatile bed of tissue, and not over a major vessel (artery or vein) that can confound the sensor and give an inaccurate SpO2 reading.[17]

SpO2 readings from forehead probes are more accurate and can detect hypoxia sooner than SpO2 obtained from digits, including in patients with hypothermia or hypotension.[14, 18, 19, 20]

 

Periprocedural Care

Equipment

The equipment used for pulse oximetry includes the following:

  • Monitoring unit

  • Probe sensor

Currently, the two basic types of pulse oximeter probes are transmission probes and reflectance probes.

Transmission probes

With transmission probes, the light emitter and sensor are placed opposite each other on pulsatile tissue such as a digit or ear. The lights used to measure tissue oxygenation are typically placed across from a detector surrounding approximately 5-10 mm of tissue that contains pulsatile blood flow, such as a fingertip or ear lobe.

Reflectance probes

With reflectance probes, the light emitter and sensor are placed side by side on a flat body surface. The detector lies adjacent to the light source on a flat surface such as the forehead. This information can be used noninvasively to help evaluate the hemodynamic status of a patient and to detect hypoxemia in various clinical settings.

Patient Preparation

Excessive debris should be removed from the area where the probe will be attached. The patient should be out of excessive light.

Positioning

The patient should be placed in a position of comfort.

 

Technique

Approach Considerations

There are several approaches to pulse oximetry. When properly attached, the ideal plethysmograph waveform shows a dicrotic notch (see image below).

Pulse oximetry with normal plethysmograph waveform Pulse oximetry with normal plethysmograph waveform.

Digit approach

This is the preferred method. The transmission probe is placed on the end of a digit, usually the finger, with the emitter on one side and the sensor on the opposite side (see image below). The digit should be resting comfortably and out of excessive light. The probe is then connected to the monitoring unit.

Portable pulse oximeter with finger probe. Portable pulse oximeter with finger probe.

Excessive debris should be removed prior to probe attachment, as well as any nail polish or artificial nails.

Ear approach

A transmission probe is placed on the end of an ear lobe with the emitter on one side and the sensor on the opposite side (see image below). The probe is then connected to the monitoring unit.

Pulse oximeter probe on ear. Pulse oximeter probe on ear.

Palm/foot approach in neonates

In neonates, in whom the digit or ear may be too small, a transmission probe may be placed over the palm or foot. The probe is then connected to the monitoring unit.

Forehead approach

A reflectance probe should be placed low across the forehead right above the eyebrows and away from a major vessel (see the images below). The patient should be resting in an inclined position. A headband around the probe and across the forehead should also be placed. The probe is then connected to the monitoring unit.

Two components for forehead blood oxygen saturatio Two components for forehead blood oxygen saturation (SpO2) sensor (Nellcorâ„¢ SpO2 Forehead Sensor).
The sensor is placed on the forehead of the patien The sensor is placed on the forehead of the patient.
The headband is placed to secure the sensor probe. The headband is placed to secure the sensor probe.